Article type: Research Article
Authors: Walburn, Frederick J.a | Blick, Edward F.b | Stein, Paul D.a
Affiliations: [a] Departments of Medicine and Surgery, Henry Ford Hospital, Detroit, MI 48202, USA | [b] School of Aerospace, Nuclear and Mechanical Engineering, University of Oklahoma, College of Engineering, Norman, Oklahoma 73019, USA
Note: [] Accepted by: Editor Y.C.B. Fung
Abstract: The purpose of this study is to explore the effect of the branch-to-trunk area ratio upon the tendency of flow to become turbulent in symmetrically branched tubes. Velocity was measured with a laser Doppler anemometer in glass tubes which branched at 70° from a 25 mm diameter trunk. The branch-to-trunk area ratio of branches studied was .4, .6, .8, 1.0, 1.2, and 1.4. For branch-to-trunk area ratios of .4 to .8, the critical Reynolds number in the trunk at which the transition to turbulence in the branch occurred was relatively constant at approximately 2100. As the branch-to-trunk area ratio increased beyond .8, the critical Reynolds number decreased until, at a ratio of 1.4, the critical Reynolds number was 1200. Since the Reynolds number of transition in the trunk was found to be 2100, it is apparent that a branch-to-trunk area ratio of 1.4 caused a prominent reduction of the critical Reynolds number in the system. A branch-to-trunk area ratio less than .8 had no effect. The velocity profiles at a branch-to trunk area ratio of .4 showed acceleration of velocity in the branch; while at an area ratio of 1.4, the velocity was shown to decelerate. This explains the tendency for flow to be unstable in branches with area ratios greater than one. These observations suggest that the branch-to-trunk area ratio usually observed in patients at the aortic bifurcation is such that laminar flow is promoted.
DOI: 10.3233/BIR-1979-16605
Journal: Biorheology, vol. 16, no. 6, pp. 411-417, 1979
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
sales@iospress.com
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
info@iospress.nl
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office info@iospress.nl
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
china@iospress.cn
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
如果您在出版方面需要帮助或有任何建, 件至: editorial@iospress.nl