You are viewing a javascript disabled version of the site. Please enable Javascript for this site to function properly.
Go to headerGo to navigationGo to searchGo to contentsGo to footer
In content section. Select this link to jump to navigation

Estimation and clinical verification of the effective and skin doses for pediatric and adult patients undergoing the cardiac interventional examination using five PMMA phantoms and TLD/ionization chamber technique

Abstract

BACKGROUND:

Effective and skin doses gain much attention since the cardiac catheterization laboratory (CCL) is a place where both patients and medical staff are exposed to X-ray or fluoroscopy environment and gain a cumulative dose during the cardiac interventional procedure.

OBJECTIVE:

These doses for pediatric and adult patients undergone cardiac interventional examination using five PMMA phantoms and thermoluminescence dosimeter (TLD)/ionization chamber technique were estimated in this work with the further clinical verification.

METHODS:

Five PMMA phantoms (10, 30, 50, 70, and 90 kg) were customized to represent baby, child, adult female, adult male, and overweight adult (by Asian complexion standards), respectively, in accordance with the ICRU-48 report. Each phantom could be disassembled into 31 plates to insert TLD chips for measuring X-ray exposed dose or assisted with an auxiliary plate to insert high-sensitivity ionization chamber for surveying low-energy fluoroscopy dose.

RESULTS:

The data acquired from five phantoms were integrated into four semi-empirical formulas, in order to fit the binary quadratic form “Dose = ABMI2+BDAP2+CBMI+ DDAP+E”. The latter linked the X-ray and fluoroscopy effective/skin doses, respectively, with a high coefficient of determination R2(from 0.888 to 0.986).

CONCLUSIONS:

The model refinement with DAP share adjustment is envisaged.

1.Introduction

Effective and skin doses of X-ray and fluoroscopy obtained by patients after the cardiac interventional examination were assessed in this work using five PMMA phantoms and thermoluminescence dosimeter (TLD)/ionization chamber technique with the further clinical verification. These doses gain much attention since the cardiac catheterization laboratory (CCL) is a place where both patients and medical staff are exposed to X-ray or fluoroscopy environment and gain a cumulative dose during the cardiac interventional procedure. The medical staff can be efficiently protected by lead aprons around the X-ray facility and protective clothing for personnel [1], whereas patients are mandatorily exposed to X-ray or fluoroscopy during the interventional examination. Several researchers have addressed this issue and applied different techniques for evaluating the cumulative doses obtained by patients of different age, weight, and size. In particular, McFadden et al. [2] established the reference level, Wu et al. [3] evaluated the exposed dose for the pediatric interventional cardiology, while Ector et al. [4] focused on the possible underestimation of accumulated doses in case of obesity/overweight patients. Chida et al. [5] addressed a surrogate measurement of the total amount of X-ray energy delivered to a patient via the dose area product (DAP) expressed in Gy × cm2 and reported a good correlation between the total entrance skin dose and DAP. However, it is quite problematic to use DAP for the maximum skin dose evaluation, since too many factors, such as field size, focus-to-image intensifier distance, and focus-to-skin distance, have to be accounted for. Hansson and Karambatsakidou [6] studied the relationships between entrance skin dose, effective dose, and DAP for patients in diagnostic and interventional cardiac procedures. This work dealt with measurements of maximum entrance skin dose and effective dose via an anthropomorphic phantom using TLDs placed both on the outside of and inserted in the phantom; and simulating a diagnostic or an interventional clinical procedure. The results [6] provided the local reference level for the patient dose to prevent skin burden. Patient mean DAP were assessed as 73, 120 and 170 Gy.cm2 and effective doses as 16, 31 and 41 mSv for diagnostic, interventional and combined procedures, respectively.

In contrast to X-ray, the fluoroscopy as frequently adopted in the cardiac interventional procedure draws less attention due to its comparatively low energy. However, the long-time exposure still creates a measurable amount of cumulative dose for patients. Thus, Chida et al. [5], Mettler et al. [7] and Koenig et al. [8] revealed the radiation injuries from fluoroscopy. Therefore, the aim of this study is to accomplish a comprehensive survey of the dose obtained from either X-ray or fluoroscopy by patients of various age and weight (from babies to overweight adult patients) via five respective PMMA phantoms and TLD/ionization chamber technique. The five phantoms are customized to simulate pediatric and adult patients of 10, 30, 50, 70, and 90 kg, respectively, according to the ICRU-48 report [9], whereas the TLD/ionization chamber was also used as a robust technique for converting the medium/low ionization energy into the exposure dose of personnel. Further, the empirical data were integrated altogether by STATISTICA developed by StatSoft, Inc. [10] to fit four semi-empirical formulas, which were defined as a binary quadratic equation to represent the effective or skin doses from X-ray or fluoroscopy by various DAP (dose area product) of facility and BMI (body mass index) of patients. The theoretical estimation of patients’ skin doses was verified by the clinical examination of 30 patients who underwent the cardiac interventional examination from May 2016 to September 2017. The discrepancies between theoretical results and clinical data and the reliability of obtained unique binary quadratic equation are discussed in detail.

2.Materials and methods

2.1Effective dose

ICRP committees have been quantifying personal radiation dose for several decades. According to the ICRP-60 report published in 1990, the protection quantity for personal dosimetry is termed the effective dose, E, and is defined as

(1)
E=ΣTωT×HT

where ωT denotes the weighting factor of particular tissue or organ, while HT represents the equivalent dose received by an tissue or organ, which is defined as

(2)
HT=Σγωγ×DT,γ

Here ωr is the weighting factor of the incident radiation, while DT,r denotes the mean dose of incident radiation of γ-type absorbed by the organ or tissue T [11]. The effective dose can be directly determined by the derivation of the equivalent dose for each organ or tissue and further multiplication of the obtained values by the corresponding weighting factors (cf. Eq. (2)). In accordance with ICRP-26, ωT values were explicitly assigned only to six organs, while five more organs receiving the next highest dose equivalents were jointly analyzed as a pseudo-organ called the “remainder” [12]. The provisions of ICRP-60 envisaged the allocation of ωT values to more organs than ICRP-26, as well as the refinement of available values of ωT, in view of more accumulated data concerning the risk of cancer in these organs due to the radiation exposure. The pseudo-organ or “remainder” was also re-defined in ICRP-60 and involved ten specific organs. The values ωT derived for particular organs were found to be independent of the equivalent dose HT delivery; therefore, the ICRP-26 definition of HE can be extended to include updated values of ICRP-60 ωT, yielding a more accurate estimate of E.

Figure 1.

(A) Five PMMA phantoms were put side by side with Rando phantom to demonstrate the relative geometrical size, (B) each phantom was assembled from 31 plates of various sizes and numbered sequentially from top to bottom as follows: Nos. 1–6 corresponded to head, 7–8 to neck, 9–21 to chest and abdomen, and 22–31 to pelvis.

(A) Five PMMA phantoms were put side by side with Rando phantom to demonstrate the relative geometrical size, (B) each phantom was assembled from 31 plates of various sizes and numbered sequentially from top to bottom as follows: Nos. 1–6 corresponded to head, 7–8 to neck, 9–21 to chest and abdomen, and 22–31 to pelvis.

2.2Five PMMA phantoms

Five PMMA phantoms (10, 30, 50, 70, and 90 kg, respectively) were customized to represent a baby, child, adult female, adult male, and overweight adult (obesity case), in accordance with ICRU-48 report, as shown in Fig. 1 [9]. Each heterogeneous phantom was assembled from 31 acrylic (PMMA) plates. In addition, the skull, ribs, spine, and pelvis were made of pure aluminum, while the lung was made of high-density polyethylene foamed cotton. Several through holes were drilled into each plate, then 3–5 TLD chips were inserted in these through holes for the dose survey, while acrylic plugs were used to fill empty ones. As shown in Fig. 1A, five PMMA phantoms were put side by side with Rando phantom to demonstrate the relative geometrical size, (B) each phantom was assembled from 31 plates of various sizes and numbered sequentially from top to bottom as follows: Nos. 1–6 corresponded to head, 7–8 to neck, 9–21 to chest and abdomen, and 22–31 to pelvis. The dimensions of five phantoms were designed according to the physical dimensions of linearity, i.e., weight W is directly proportional to the cubed linear dimension L(WL3). Therefore, the body mass index (BMI) was 13.52, 16.46, 21.64, 24.22, and 27.78 kg m-2 for the phantoms of 10, 30, 50, 70, and 90 kg, respectively. Noteworthy is that not only the arrangement of each plate but also the structure of five phantoms were identical to ensure the consistence and varied only by size. Thus, different exposed doses obtained via either X-ray or fluoroscopy can be only correlated to phantoms of different size.

2.3TLD/ionization chamber setup

One hundred five standard TLD-100 chips consisting of Lithium Fluoride (LiF: Mg, Ti; 3.2 × 3.2 × 0.89 mm3) were obtained from Thermo Fisher Scientific Inc. (formerly Harshaw, Bicron). The TLDs were randomly categorized into 35 packs, each consisting of three TLDs. The TLD assigned dose was calibrated using the Advanced Markus ionization chamber, namely Victoreen Model 4000 with a parallel plate ionization chamber. The irradiated TLD reading was obtained using a Mikro Lab RA94 TLD reader/analyzer and by annealing in furnaces (Barnstead Int. Co., model 47900) coupled with an oven/incubator (model 19200) for 400C 1 h and 100 C 2 h. Furthermore, each TLD was cooled for at least 24 hours before the next exposure, in order to efficiently suppress the residual dose. The TLD signal fading and self-absorption of TLD light can be treated as an internal interfering factor and given a negligible contribution, insofar as the exposed TLDs were always read in the same period of cooling time.

An additional high-sensitivity pencil-type ionization chamber (Victoreen, model 6000–100, 3.2 cc) was preset for measuring the fluoroscopy, since TLD had comparatively low detecting efficiency for fluoroscopy. In doing so, an auxiliary PMMA plate was specially made with 3–5 through tunnels. The pencil-type ionization chamber could be inserted into any tunnel to survey the dose, whereas acrylic sticks of the same size were inserted into the remaining 2–4 through tunnels to maintain the plate integrity, yet, the auxiliary plate could be inserted into any layer to replace the original one of the phantom. The pencil-type ionization chamber had a high sensitivity to measure the low-energy fluoroscopy and maintain the dose-response linearity, its only drawback being the inconvenience of its application and larger time required for the dose measurement. The pencil-type ionization chamber could record a single fluoroscopy dose at a time to obtain a single dose position inside the phantom. In contrast, 105 TLDs could be fully inserted into the phantom and then exposed only once to acquire the full empirical data for a single specific task.

Table 1

The tests on X-ray or fluoroscopy of phantoms were verified using real patients undergone the cardiac interventional examination. Here 10- and 30-kg phantoms corresponded to baby or child, respectively, from the pediatric viewpoint. Similarly, 50-, 70-, and 90-kg phantoms were used to simulate the adult female, male, and overweight adult, respectively

PediatricAdult
Phantom10 kg30 kg50 kg70 kg90 kg
X-ray
 kVp6163666770
 mA135267375430577
Fluoroscopy
 kVp6065686875
 mA3.05.78.38.312.4

The phantom was assembled with TLD chips or pencil-type ionization chamber inside, then exposed to the bi-plane X-ray facility that was specifically designed for the cardiac interventional examination (Philips Integris Allura 9 Biplane system) located at the Cardiac Catheterization Laboratory, Taichung Veteran General Hospital, Taiwan (CCL, TVGH). The X-ray and fluoroscopy settings for each phantom were then applied to real patients that underwent the cardiac interventional examination as listed in Table 1. Here 10- and 30-kg phantoms corresponded to baby or child, respectively, from the pediatric viewpoint. Similarly, 50-, 70-, and 90-kg phantoms were used to simulate the adult female, male, and overweight adult, respectively. The recorded DAP could be manipulated by changing the exposure time. Table 2 implies the precise arrangement of well-packed TLDs inside the phantom for X-ray exposure. Each TLD pack had three chips sealed in a PE bag, which were then inserted into specific through holes for the exposure. Also, the tissue weighting factor (cf. Eq. (1), ωT) was also listed for reference; the factor was normalized from the original ICRP-60 report to ensure its unity. Figure 2 reveals part of the data acquisition process inside the CCL. As seen in Fig. 2A, a 70 kg phantom was placed between two X-ray emitters, and the focal spots aiming at the back of phantom, (B) a 10 kg phantom, the photo was taken from the opposite side of (A), (C) five auxiliary plates with 3–5 through tunnels to insert the pencil-type ionization chamber for surveying the fluoroscopy dose. The size of an auxiliary plate equaled by the geometrical size to the respective phantom, which allowed one to insert it into any layers to replace the original one, and (D) a 90 kg phantom with auxiliary plate replacing the original 14th plate to measure the fluoroscopy dose; as is seen, the acrylic stick was extracted from the tunnel, in order to put in the pencil-type ionization chamber from another side, whereas other four through tunnels were filled with sticks to maintain the plate integrity.

Table 2

Implies the precise arrangement of well-packed TLDs inside the phantom for X-ray exposure. Each TLD pack had three chips sealed in a PE bag, which were then inserted into specific through holes for the exposure. Also, the tissue weighting factor (cf. Eq. (1), ωT) was also listed for reference; the factor was normalized from the original ICRP-60 report to ensure its unity

Organ or tissueICRP-60 ωTPhantom plate no.No. of TLD packPrecise position
Thyroid0.0582
Esophagus0.0592
Breast0.05152Left side
152Right side
Lung0.12112Left side
142Right side
Heart0.05102Left atrium
112Right atrium
142Right ventricle
152Left ventricle
Liver0.05212
Stomach0.12202
Bone surface0.0191Clavicle
91Thoracic vertebra
111
151
181
111Left side
171Right side
161Sternum
Skin0.01111X-ray focal spot on patient’s back
121
131
Colon0.1224N/A
Bladder0.0528N/A
Gonads0.2031N/A
Sum 1.00 35

Figure 2.

(A) A 70 kg phantom was placed between two X-ray emitters, and the focal spots aiming at the back of phantom, (B) a 10 kg phantom, the photo was taken from the opposite side of (A), (C) five auxiliary plate with 3-5 through holes to insert the pencil-type ionization chamber for surveying the fluoroscopy dose. The size of an auxiliary plate equaled by the geometrical size to the respective phantom, which allowed one to insert it into any layers to replace the original one, and (D) a 90 kg phantom with auxiliary plate replacing the original 14th plate to measure the fluoroscopy dose; as is seen, the acrylic plug was extracted from the hole, in order to put in the pencil-type ionization chamber from another side, whereas other four through holes were filled with plugs to maintain the plate integrity.

(A) A 70 kg phantom was placed between two X-ray emitters, and the focal spots aiming at the back of phantom, (B) a 10 kg phantom, the photo was taken from the opposite side of (A), (C) five auxiliary plate with 3-5 through holes to insert the pencil-type ionization chamber for surveying the fluoroscopy dose. The size of an auxiliary plate equaled by the geometrical size to the respective phantom, which allowed one to insert it into any layers to replace the original one, and (D) a 90 kg phantom with auxiliary plate replacing the original 14th plate to measure the fluoroscopy dose; as is seen, the acrylic plug was extracted from the hole, in order to put in the pencil-type ionization chamber from another side, whereas other four through holes were filled with plugs to maintain the plate integrity.

3.Results

3.1Effective and skin doses

Table 3 lists the relevant data derived in this work. The reported data for each specific case were averaged from three independent measurements. The TLDs were measured and then averaged to represent the particular assigned organ or tissue. The data were categorized by BMI and exposed time Er [sec] (cf. Table 3), then the time was converted to different DAP of X-ray or fluoroscopy, in compliance with a preliminary survey that was calibrated using both BMI and exposed time [13]. Figure 3 depicts four correlations between DAP and effective/skin dose for five different phantom weights. Thus, a thorough survey of effective or skin doses for five phantoms exposed to X-ray or fluoroscopy was conducted. The high linearity or consistence among five correlation plots revealed a quite robust fit with a negligible systematic error.

Table 3

The reported data for each specific case were averaged from three independent measurements. The TLDs were measured and then averaged to obtain the values representing the assigned organ or tissue. The data were categorized by BMI and exposed time

PMMAEr X Y1 Z1 (X-ray) Y2 Z2(Fluor)

Phantom
(kg)

time (sec)

BMI
(kg/m)2

DAP
(mGy cm)2

Effective
dose (μSv)

Skin dose
(μSv)

DAP
(mGy cm2)

Effective
dose (μSv)

Skin dose
(μSv)

107013.52260622611211512128220
1014013.52527542718363075191329
1021013.52794960525294547258444
1028013.521055581534916059321553
1042013.5215830129479909134452780
307016.46923831633752727152917
3014016.4618336444369054022702205
3021016.4627579669578181133713104
3028016.463681710298978108404584278
3042016.4655153144112071162427056470
507021.6435325479168739781461118
5014021.6470773811477778902712369
5021021.6410623310845715118863793438
5028021.64141558170410104158645264638
5042021.64212331237812187237537726951
707024.2244415325223751491561467
7014024.22883149224997102382773249
7021024.221323519416200153184024979
7028024.22176766163812118204675276766
7042024.222650802189185673070577610302
907027.7856191344143260261432659
9014027.781115866243436121432534892
9021027.7816775111774345181103607133
9028027.7822394218679780241364699352
9042027.783355282593126703627968813848

Figure 3.

Four correlations between DAP and dose using five phantoms of different weight. Thus, a thorough survey of effective or skin doses for five phantoms exposed to X-ray or fluoroscopy was accomplished in this work. (A) Effective dose from X-ray exposure, (B) skin dose from X-ray exposure, (C) effective dose from fluoroscopy, and (D) skin dose from fluoroscopy.

Four correlations between DAP and dose using five phantoms of different weight. Thus, a thorough survey of effective or skin doses for five phantoms exposed to X-ray or fluoroscopy was accomplished in this work. (A) Effective dose from X-ray exposure, (B) skin dose from X-ray exposure, (C) effective dose from fluoroscopy, and (D) skin dose from fluoroscopy.

3.2Error treatment

The errors associated with the derived effective or skin doses for various X-ray/fluoroscopy exposure arrangements were calculated as the square root of the sum of squared individual errors, Δi, as listed in Table 4. The uncertainty for ωT value was set to 5%, since the weighting factor was normalized in this specific measurement. The error for X-ray power fluctuation was based on the monthly clinical quality assurance (QA) at Taichung Veteran Hospital, while the internal normalization errors for TLDs were quoted from the preliminary survey of each specific TLD, as listed in Table 4 [14, 15]. The uncertainty resulting from non-tissue equivalence effect in the acrylic phantom was set to 5%, since the heterogeneous phantom was manufactured from PMMA, aluminum, and high-density polyethylene foamed cotton. Eventually, the total error as indicated in this study was mainly related to the statistical error in counting, which could be efficiently suppressed by repeated measurements. Therefore, all reported data were averaged from three independent trials, and the maximum counting statistical and total errors amounted to 6.2% and 15.4%, respectively.

Table 4

The errors associated with the derived effective or skin doses for various X-ray/fluoroscopy exposure arrangements. The total error was calculated as the square root of the sum of squared individual errors

SourceError (%)
Systematic
ωT (tissue or organ weighting factor)5%
 Non-tissue equivalent effect5%
 Fluctuation of X-ray or Fluoroscopy3%
 Internal normalization of TLD chip3% 8%
 Pencil type ion chamber dose conversion1.1% 2.5%
Random
 TLD chips reading statistics2.7% 5.1%
 Pencil type ion chamber repeat statistics3.8% 6.2%
Δ tot 8.9% 15.4%

4.Discussion

4.1Optimal-fitting of a semi-empirical formula via STATISTICA program

The correlation of DAP versus dose for various phantom weights (reduced to BMI) can be expressed by a semi-empirical formula defined as a binary quadratic equation to predict the exposure dose (effective or skin one) for patients that underwent the cardiac interventional examination (via X-ray or fluoroscopy) in CCL. In doing so, the coefficients of the semi-empirical formula can be derived with the STATISTICA program [10]. The correlations among the variables are determined and defined as nonlinear models, nonlinear estimations, and user-specified regressions with customized loss functions to perform the numerical analysis using the normalized data from the real measurement. The predicted effective or skin doses are the expectation values of the computational results. Therefore, five independent groups of 25 individual datasets each [5 × 5 = 25] were incorporated into the model to optimize the compromised solution of the predicted dose. Also, five terms, including one constant, were used in the binary quadratic equation to reveal the best correlation among the variables as listed below:

(3)
Dose=AX2+BY2+CX+DY+E

Table 5

The derived coefficients AE for the four semi-empirical formulas and the coefficient of determination, r2

SourceDoseDose = ABMI2+ BDAP2+ CBMI+ DDAP+ E
ABCDE r2
X-rayEffective-0.4060.000-67.160.0141569.40.952
Skin-37.0500.000897.20.087-1968.80.888
FluoroscopyEffective-6.36600.000196.50.039-1278.90.977
Skin-9.69600.000711.40.293-10019.00.986

Figure 4.

Plot of the expectation values predicted via the semi-empirical formulas, which was automatically constructed by the STATISTICA default feature. (A) effective dose from X-ray exposure, (B) skin dose from X-ray exposure, (C) effective dose from fluoroscopy, and (D) skin dose from fluoroscopy.

Plot of the expectation values predicted via the semi-empirical formulas, which was automatically constructed by the STATISTICA default feature. (A) effective dose from X-ray exposure, (B) skin dose from X-ray exposure, (C) effective dose from fluoroscopy, and (D) skin dose from fluoroscopy.

where Dose, X, and Y are the expectation values of effective or skin dose from X-ray or fluoroscopy, BMI, and DAP, respectively, whereas terms AE are the derived coefficients of the formula. The loss function was defined as the total deviation between the predicted and observed doses for all 25 cases, and thus, a small loss function is always preferable in the theoretical computation. Table 5 lists the derived coefficients AE for the four binary quadratic equations and the coefficient of determination R2, whereas the respective plot for expectation values is depicted in Fig. 4. The latter was automatically plotted by the STATISTICA default feature, and the smooth surface of the expectation values’ domain indicates the lack of conflicts among datasets and the absence of any systematic errors within the computational process. Thus, the coefficient of determination R2 reaches 0.888 0.986, which implies a high consistence and reliable estimation for the four kinds of exposed doses in CCL. Eventually, the effective or skin doses can be easily derived by including the patient’s BMI and recorded DAP of X-ray or fluoroscopy. Also, the quadratic coefficient of DAP reaches 0.0 in all four equations (cf. Table 5). This prevents too large DAP values to confuse the estimation, since DAP values may vary by several thousand units [mGycm-2], whereas the BMI ranges only by several ten units [kgm2]. However, this biased estimation can be suppressed by normalizing the respective variable to the same level and reducing its fluctuation to the range from -1.0 to +1.0 [16, 17].

4.2Clinical verification using 30 cardiac interventional patients

The obtained binary quadratic equations were applied to predict the exposed dose for 30 patients who underwent the cardiac interventional examination. This was envisaged to verify the theoretical estimation accuracy. In doing so, thirty patients were asked to place three TLDs on their backs within the X-ray focal spot aiming region during the cardiac examination. The TLD chips had to be put along the cardiac edge to avoid any false imaging being made. The TLD chips were removed immediately after the examination and followed the same process of TLD reading to collect the clinical data. Table 6 lists the derived data and their comparison with the theoretical estimation. As seen in Table 6, these results are widely fluctuated. Only 12 out of 30 cases have a disagreement lower than 100%, whereas others exhibit even worse correlation. The possible reasons of such deviations might be (A) the focal spot aiming of the X-ray is moving according to instant response from the cardiac interventional examination. In contrast, the equation is calculated according to fixed DAP that is exposed by a steady focal spot aiming, thus, the prediction might under- or overestimate the dose on the basis of inappropriate DAP index, (B) the facility can show only combined DAP index rather than individual DAP by X-ray or fluoroscopy, respectively. Thus, the DAP is roughly divided into fifty-fifty contributions of fluoroscopy or X-ray. The assumption is concluded from a long-term in-situ statistics in the CCL from May 2016 to September 2017 and suggested a “fifty-fifty” rule of thumb. The X-ray has higher specific DAP than fluoroscopy (for 70 kg phantom, X-ray and fluoroscopy exhibited approx. 630 and 72 DAP/s, respectively; whereas the time sharing between X-ray and fluoroscopy reached the ratio of 1:9 in the regular CCL diagnosis. Thus, DAP fulfills the “fifty-fifty” rule in reality. In the routine examination, the CCL staff needs to step on the foot pedal of fluoroscopy for acquiring the tentative image and then refine the preset parameters for ensuring a precise image (since the machine operates only when there is pressure on the switch); whereas the X-ray is used to ascertain the cardiac interventional process for the diagnosis or to store the printed imaging for the follow-up study. In some special cases, the fluoroscopy is continuously adopted for the benefit of low-exposure dose, and then the DAP ratio changes to “forty-sixty”. Therefore, from the statistical viewpoint, the exact breakdown of DAP between fluoroscopy and X-ray is a complicated and fuzzy process. Yet, from the clinical viewpoint, the “fifty-fifty” option is always a good compromise. For example, if a 70-kg standard male patient receives 60,000 DAP during the cardiac interventional examination, it is roughly estimated that each source shares 30,000 DAP. Then, the derived values for X-ray effective/skin dose and fluoroscopy effective/skin dose are assessed as 125, 638, 916, and 10313 μSv, respectively. If the DAP breakdown is changed to “forty-sixty” ratio, then the respective doses delivered to the patient will be 41, 116, 1150, and 12071 μSv (cf. Tabl 5). However, these predictions are based on the scenario that the focal spot experiences no shift during the examination process. Otherwise, the skin dose will be distributed along a larger area and, thus, sharply reduced.

Table 6

The derived data according to the semi-empirical formula and compared to clinical data obtained. Only 12 out of 30 cases exhibit a good fit, while the remaining ones disagree by more than 100%

X-rayFluoroscopyEffective doseSkin dose
No.

BMI
(kg/m2)

Er
Time
(sec)

Time

DAP
(mGy.
cm2)

Time

DAP
(mGy.
cm2)

Predicted
(μSv)

Predicted
[A]
(μSv)

Clinical
[B]
(μSv)

AT (%)
[A]-[B]/[B]
100%

124.57313228518211328472067596938821485360653
220.7040237153803652020910348614593445
326.3116861531131091533123237427246131167972-73
420.9035432139083221808490677811175562
524.0554529964934513429361032-9
619.05424125138182938523383754-38
723.1215014760413690173053535180096
827.221254114899911140963033108345399027283
921.7327025117592451470462162651500318
1021.101746159703901587906143936351903515901
1126.439909067027900728462350258296955271
1220.644984518900453249131260103551921439
1322.9494286468838565593922932128742788-50
1427.681188108880631080934572896330101729891
1524.1366636886042393181177576104
1624.171681594231531081733035945321-32
1721.5128826122182621542768966121406370
1820.2446843167994252265112099613501092
1924.89787466471525142010411329-22
2020.432282184072071121164854061179359
2123.81450412440840928315909103642163379
2223.53450412376140927803936103233142229
2323.1421019106681911264140048661290277
2421.691921783231751042542147112100124
2518.9113813401112559325983810434778
2624.1523421131012131504843951852352120
2717.0222821441220780488994606721539
2827.303363124251305259135546894583218
2923.431801694121641104835341181873120
3019.43504461599445822732131597691010867

5.Conclusions

The effective and skin dose for patients undergone cardiac interventional examination were assessed in this work using five PMMA phantoms and TLD/ionization chamber technique with the further clinical verification. Five PMMA phantoms (10, 30, 50, 70, and 90 kg) were customized to represent baby, child, adult female, adult male, and overweight adult, in compliance with ICRU-48 report. The data obtained from five phantoms were integrated into four semi-empirical formulas to describe the X-ray and fluoroscopy effective/skin doses with a high coefficient of determination. The predicted dose was later verified by the clinical survey, where 40% (12 cases of 30) revealed a good fit. Possible reasons for high deviations (over 100%) in the remaining 60% of cases could be attributed to the deviated focal spot aiming at patient’s back during the diagnosis or erroneous estimation of DAP shares corresponding to X-ray or fluoroscopy. These factors, as well as other ones reported by other authors (e.g., variations in the field size, focus-to-image intensifier distance, and focus-to-skin distance), have to be accounted for in further research efforts.

Acknowledgments

The authors would like to thank the National Defense-Medical Affairs Bureau (contract no. ND 106-A11) and the Ministry of Science and Technology of the Republic of China (contract no. MOST 106-2221-E-166-002) for the financial support of this research.

Conflict of interest

None to report.

References

[1] 

Pan LF, Kittipayak S, Yen SL, et al. Evaluation of the occupational X-rays dose of the medical staff in a cardiac catheterization laboratory using an acrylic phantom and semiconductor dosimeter, Hellenic Journal of Nuclear Medicine. (2016) ; 19: (2): 140-146.

[2] 

McFadden S, Hughes C, Dhelft CI, et al. The establishment of local diagnostic reference levels for pediatric interventional cardiology, Radiography. (2013) ; 19: : 295-310.

[3] 

Wu JR, Huang TY, Wu DK, et al. Radiation exposure of pediatric patients and physicians during cardiac catheterization and balloon pulmonary valvuloplasty, American Journal of Cardiology. (1991) ; 68: (2): 221-225.

[4] 

Ector J, Dragusin O, Adriaenssens B, et al. Obesity is a major determinant of radiation dose in patients undergoing pulmonary vein isolation for atrial fibrillation, Journal of American College of Cardiology. (2007) ; 50: (3): 234-242.

[5] 

Chida K, Saito H, Otani H, et al. Relationship between fluoroscopic time, dose-area product, body weight, and maximum radiation skin dose in cardiac interventional procedures, American Journal of Roentgenology. (2006) ; 186: (3): 774-778.

[6] 

Hansson B, Karambatsakidou A. Relationships between entrance skin dose, effective dose and dose area product for patients in diagnostic and interventional cardiac procedures. Radiation Protection Dosimetry. (2000) ; 90: (1-2): 141-144.

[7] 

Mettler FA, Koenig TR, Wagner LK, et al. Radiation injuries after fluoroscopic procedures. Seminars in Ultrasound, CT, and MRI. (2002) ; 23: (5): 428-442.

[8] 

Koenig TR, Wolff D, Mettler FA, et al. Skin injuries from Fluoroscopically guided procedures Part 1, characteristics of radiation injury. American Journal of Roentgenology. (2001) ; 177: : 3-11.

[9] 

ICRU-48. Phantoms and computational models in therapy, diagnosis and protection, International commission on radiation units and measurement. (1992) ; https://academic.oup.com/jicru/issue/os25/1.

[10] 

STATISTICA version 10, 2017. Available at: http://www.statsoft.com/Products/STATISTICA-Features/Version-10.

[11] 

ICRP-60. Recommendations of the international commission on radiological protection Technical Report, International commission on radiation protection, (1991) ; http://www.icrp.org/publication.asp?id=icrp%20publication%2060.

[12] 

ICRP-26. Recommendations of the international commission on radiological protection Technical Report, International commission on radiation protection, (1977) ; http://www.icrp.org/publication.asp?id=ICRP%20Publication%2026.

[13] 

Chang ZK. To evaluate the effective dose for patient undergone cardiac angiography using five PMMA phantoms, Mater thesis, Dept. of medical imaging and radiological sciences, Central Taiwan University of Science and Technology. Taichung, Taiwan, (2013) .

[14] 

Chen CY, Yang KC, Pan LK. Bubble technique for evaluating effective dose of diagnostic X-rays: a feasibility study, J. Radiation Research, (2009) ; 50: : 449-456.

[15] 

Liu MT, Huang SS, Yeh DM, et al. Distribution of spatial photoneutrons inside a 70 kg water phantom via neutron activation analysis, Applied Radiation Isotopes. (2010) ; 68: : 1816-1821.

[16] 

Pan LF, Davaa O, Chen CY, et al. Quantitative evaluation of contrast-induced-nephropathy in vascular post-angiography patients: feasibility study of a semi-empirical model, Bio-Medical Materials and Engineering. (2015) ; 26: : s851-860.

[17] 

Pan LF, Chiu SW, Xiao MF, et al. Revised inverse problem algorithm-based prediction of coronary artery stenosis readings from the clinical data of patients with coronary heart diseases, Computer Assisted Surgery, (2017) . http//dx.doi.org/10.1080/24699322.2017.1379149.