Searching for just a few words should be enough to get started. If you need to make more complex queries, use the tips below to guide you.
Article type: Research Article
Authors: Uchiyama, Akinoria; * | Nishimura, Masajib | Amata, Mitsuyukic | Mashimo, Takashia | Fujino, Yujia
Affiliations: [a] Osaka University Hospital Intensive Care Unit, Osaka, Japan | [b] Department of Emergency and Critical Care Medicine, The University of Tokushima Graduate School, Tokushima, Japan | [c] Department of Anesthesiology, Osaka University Medical School, Osaka, Japan
Correspondence: [*] Address for correspondence: Akinori Uchiyama, Osaka University Hospital Intensive Care Unit, 2-15 Yamadaoka, Suita, Osaka Prefecture 565-0871, Japan. Tel.: +81 6 6879 5824; Fax: +81 6 6879 582; E-mail: auchiyama@hp-icu.med.osaka-u.ac.jp.
Abstract: The aim of the study was to assess a new expiratory support (ES) system in resolving air trapping in the lungs during mechanical ventilation. The ES system consisted of a cylinder and two valves that were connected to the ventilatory circuit. The movements of them were synchronized with the ventilator. The cylinder aspirated gas during expiration. We compared the effects of the ES on air trapping between a narrower and an ordinary size endotracheal tube (ETT) (internal diameter (ID): 5 and 8 mm). In addition, we compared the effects of the ES on air trapping between normal and high airway resistance of the lungs (5 and 20 cm H2O/L/s). The ventilatory settings were as follows: volume controlled ventilation with constant inspiratory flow rate; tidal volume, 0.5 L; inspiratory time, 1.0 s; expiratory time, 1.0, 1.5, or 2.0 s; and PEEP, 0 cm H2O. The ES normalized the end-expiratory alveolar pressure of the 5 mm ID ETT at a level similar to that of the 8 mm ID ETT. The ES also resolved the air trapping induced by the high airway resistance of the lungs. In conclusion, the ES system resolved the air trapping associated with a narrow endotracheal tube and high airway resistance of the lungs.
Keywords: Mechanical ventilation, endotracheal tube, expiratory support, air trapping
DOI: 10.3233/THC-2007-15305
Journal: Technology and Health Care, vol. 15, no. 3, pp. 213-220, 2007
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
sales@iospress.com
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
info@iospress.nl
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office info@iospress.nl
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
china@iospress.cn
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
如果您在出版方面需要帮助或有任何建, 件至: editorial@iospress.nl