Article type: Research Article
Authors: Kweon, Dae Cheol; *
Affiliations: Department of Radiological Science, Shinhan University, Uijeongbu, Republic of Korea
Correspondence: [*] Correspondence author: Dae Cheol Kweon, PhD, Associate Professor, Department of Radiological Science, Shinhan University, 95 Hoam-ro, Uijeingbu-si, Gyeonggi-do, 11644 Republic of Korea. Tel.: +82 31 870 3411; Fax: +82 31 870 3419; E-mail: dckweon@shinhan.ac.kr.
Abstract: BACKGROUND:Due to the wide application of the cone beam computed tomography (CBCT) in clinical practice, it is important to assess radiation dose of CBCT more accurately and efficiently in different clinical applications. OBJECTIVE:This study aims to calculate effective and absorbed doses in CBCT measured in an anthropomorphic phantom using computer-based Monte Carlo (PCXMC) software, and to conduct comparative evaluations of MOSFET (metal– oxide– semiconductor field-effect transistor) and radiophotoluminescence glass dosimeters (RPLGD). METHODS:Effective and absorbed organ doses are compared with those obtained using MOSFET and RPLGD dosimetry in an anthropomorphic phantom given the same exposure settings. Effective and absorbed organ doses from CBCT during scout and main projections are calculated using PCXMC and PCXMCRotation software, respectively. RESULTS:The mean effective dose from CBCT calculated using PCXMC software is 233.8μSv, while the doses calculated using dosimetry (MOSFET and RPLGD) are 266.67μSv and 268.78μSv, respectively. The X-ray source variation is 0.79%. The prescription limits based on the Friedman test for MOSFET and RPLGD pre-points (i.e., in an analytical analysis of diagnostic names in CBCT) are not statistically significant. The calculated correlation coefficient between MOSFET- and RPLGD-derived absorbed dose values with respect to a field of view CBCT parameter of 17×13.5 mm is r = 0.8623. CONCLUSIONS:The study demonstrates that the PCXMC software may be used as an alternative to MOSFET and RPLGD dosimetry for effective and absorbed organ dose estimation in CBCT conducted with a large FOV in an anthropomorphic phantom.
Keywords: Absorbed dose, cone beam CT, radiation dose, MOSFET, RPLGD
DOI: 10.3233/XST-230026
Journal: Journal of X-Ray Science and Technology, vol. 31, no. 4, pp. 837-852, 2023
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
sales@iospress.com
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
info@iospress.nl
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office info@iospress.nl
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
china@iospress.cn
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
如果您在出版方面需要帮助或有任何建, 件至: editorial@iospress.nl