Searching for just a few words should be enough to get started. If you need to make more complex queries, use the tips below to guide you.
Article type: Research Article
Authors: Isa, I.N.Ca | Rahmat, S.M.S.b; c; * | Dom, S.M.c | Kayun, Z.d | Karim, M.K.Ad; e; *
Affiliations: [a] Diagnostic Imaging & Radiotherapy Program, Faculty of Health Sciences, Universiti Kebangsaan Malaysia, Kuala Lumpur, Malaysia | [b] Department of Radiology, National Cancer Institute, Putrajaya, Malaysia | [c] Faculty of Health Sciences, Universiti Teknologi MARA, Puncak Alam, Selangor, Malaysia | [d] Department of Physics, Faculty of Science, Universiti Putra Malaysia, Serdang, Selangor, Malaysia | [e] Center for Diagnostic Nuclear Imaging, Faculty of Medicine, Universiti Putra Malaysia, Serdang, Selangor, Malaysia
Correspondence: [*] Corresponding authors: M.K.A. Karim, Department of Physics, Faculty of Science, Universiti Putra Malaysia 43400 Serdang, Selangor, Malaysia. E-mail: khalis.karim@gmail.com and S.M.S. Rahmat, Department of Radiology, National Cancer Institute, Putrajaya, Malaysia, E-mail: smshaffiq@yahoo.com.
Abstract: There are several factors that may contribute to the increase in radiation dose of CT including the use of unoptimized protocols and improper scanning technique. In this study, we aim to determine significant impact on radiation dose as a result of mis-centering during CT head examination. The scanning was performed by using Toshiba Aquilion 64 slices multi-detector CT (MDCT) scanner and dose were measured by using calibrated ionization chamber. Two scanning protocols of routine CT head; 120 kVp/ 180 mAs and 100 kVp/ 142 mAs were used represent standard and low dose, respectively. As reference measurement, the dose was first measured on standard cylindrical polymethyl methacrylate (PMMA) phantom that positioned at 104 cm from the floor (reference isocenter). The positions then were varied to simulate mis-centering by 5 cm from isocenter, superiorly and inferiorly at 109 cm, 114 cm, 119 cm, 124 cm and 99 cm, 94 cm, 89 cm, 84 cm, respectively. Scanning parameter and dose information from the console were recorded for the radiation effective dose (E) measurement. The highest mean CTDIvol value for MCS and MCI were 105.06 mGy (at +10 cm) and 105.51 mGy (at – 10 cm), respectively which differed significantly (p < 0.05) as compared to the isocenter. There were large significant different (p < 0.05) of mean Dose Length Product (DLP) recorded between isocenter to the MCS (85.8 mGy.cm) and MCI (93.1 mGy.cm). As the low dose protocol implemented, the volume CTDI (CTDIvol) were significantly increase (p < 0.05) for MCS (at +10 cm) and MCI (at – 10 cm) when compared to the isocenter. The phantom study revealed a noticeable different in radiation dose between isocenter and experimental groups due to degradation of the bowtie filter performance. It is anticipated that these noteworthy findings may emphasize the importance of accurate patient centering at the isocenter of CT gantry, so that CT optimization practice can be achieved.
Keywords: Mis-centering position, CT head, radiation dose, CT dose index
DOI: 10.3233/XST-190491
Journal: Journal of X-Ray Science and Technology, vol. 27, no. 4, pp. 631-639, 2019
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
sales@iospress.com
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
info@iospress.nl
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office info@iospress.nl
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
china@iospress.cn
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
如果您在出版方面需要帮助或有任何建, 件至: editorial@iospress.nl