Searching for just a few words should be enough to get started. If you need to make more complex queries, use the tips below to guide you.
Article type: Research Article
Authors: Lu, Boa | Deng, Hepingb | Chen, Baichenga; * | Zhao, Jianc
Affiliations: [a] Department of Orthopaedics, The Third Hospital of Hebei Medical University, Shijiazhuang, Hebei, P.R. China | [b] Department of Ultrasonography, The Third Hospital of Hebei Medical University, Shijiazhuang, Hebei, P.R. China | [c] Department of Radiology, The Third Hospital of Hebei Medical University, Shijiazhuang, Hebei, P.R. China
Correspondence: [*] Corresponding author: Baicheng Chen, Department of Orthopaedics, The Third Hospital of Hebei Medical University, No. 139, Ziqiang Road, Shijiazhuang, 050000, Hebei, P.R. China. Tel.: +86 13313012818; E-mail: chenbaichengmed@163.com.
Abstract: OBJECTIVE:To investigate and evaluate the accuracy of ultrasound (US) imaging for measurement of ischiofemoral space (IFS) compared with magnetic resonance imaging (MRI). METHODS:Twenty-five hips of 16 patients with hip pain and ipsilateral quadratus femoris muscle (QFM) edema were recruited to the IFI group, and 19 hips of 19 patients without hip pain and QFM edema were recruited as the control group. IFS of both groups was measured by US and MRI. The correlations and differences between US and MRI measurements were then assessed. Last, the receiver operating characteristic (ROC) data analysis was performed. RESULTS:The US results revealed a decreased IFS in IFI hips compared with controls (P < 0.001), suggesting that US were valuable in identifying IFI. IFS measured by US and MRI showed positive correlations in both the IFI group (r = 0.409, P = 0.042) and control group (r = 0.575, P = 0.01). There were no statistically significant differences between IFS measured by US and MRI in the control group (P = 0.657), while IFS measurements in the IFI group performed with US were significantly greater than those with MRI (P < 0.001). ROC curve analysis revealed that the cutoff value of IFS measured with US was 2.14 cm, with a sensitivity of 92.0% and specificity of 68.4%, while measured by MRI was 1.87 cm, with a sensitivity of 96.0% and specificity of 84.2%. CONCLUSIONS:IFS measurements obtained with US are very similar to those obtained with MRI. Therefore, US provides a low-cost and easy-to-operate alternative imaging modality to measure IFS.
Keywords: Ishiofemoral, impingement, ultrasound, magnetic resonance images
DOI: 10.3233/XST-180482
Journal: Journal of X-Ray Science and Technology, vol. 27, no. 4, pp. 605-614, 2019
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
sales@iospress.com
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
info@iospress.nl
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office info@iospress.nl
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
china@iospress.cn
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
如果您在出版方面需要帮助或有任何建, 件至: editorial@iospress.nl