Article type: Research Article
Authors: Tang, Shaojie | Mou, Xuanqin | Yang, Ying | Xu, Qiong | Yu, Hengyong
Affiliations: Institute of Image Processing & Pattern Recognition, Xi'an Jiaotong University, Xi'an, Shaanxi, 710049, P.R. China | Biomedical Imaging Division, VT-WFU School of Biomedical Engineering and Science, Virginia Tech., Blacksburg, VA 240601, USA
Note: [] Corresponding author. Prof. Xuanqin Mou, Institute of Image Processing & Pattern Recognition, Xi'an Jiaotong University, Xi'an, Shaanxi, 710049, P.R. China. E-mail: xqmou@mail.xjtu.edu.cn
Abstract: The polychromatic spectrum of photons emitted by an X-ray tube in medical diagnostic CT system brings beam hardening artifact. Although many correction methods have been developed, it is still the major artifact in clinical CT imaging. Motivated by CT reconstruction researches benefiting from mathematical phantoms, one so-called physical mathematical phantom and its projection simulation are presented in this paper, which are used in comprehensive evaluation of beam hardening corrections. Due to the inherent merits as compared to discrete and stochastic simulations, an analytic method is proposed to simulate the polychromatic projections of FORBILD head phantom. Firstly, CT-numbers are mapped to material compositions under the precondition sub-regions represent different human tissues. Second, boundary parameters of sub-regions are determined by the presented algorithms. Finally, projections are calculated according to X-ray energy spectrum. For validating the usefulness of our simulation, several typical correction methods are analyzed using simulated projections. The performance of every correction is exhibited distinctly. The analysis results show that there are still the spaces to improve those correction methods.
Keywords: Polychromatic spectrum, beam hardening, projection simulation, FORBILD
Journal: Journal of X-Ray Science and Technology, vol. 16, no. 2, pp. 95-117, 2008
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
sales@iospress.com
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
info@iospress.nl
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office info@iospress.nl
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
china@iospress.cn
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
如果您在出版方面需要帮助或有任何建, 件至: editorial@iospress.nl