Abstract: BACKGROUND: In spite of widespread use of nasal CPAP there are comparatively few studies to guide the choice of nasal prongs. OBJECTIVES: To determine whether the Fisher & Paykel Healthcare (FPH) neonatal continuous positive airway pressure (CPAP) interface was effective in providing bubble CPAP when compared to the Hudson prong interface. METHODS: The study was a randomized cross-over study of twenty newborn infants 500 g or more requiring CPAP for respiratory support at birth. Infants were randomized to either the Fisher & Paykel Healthcare or Hudson CPAP interface for twenty four hours. Crossover between interfaces occurred after subsequent twenty four hour periods. The primary outcome was the provision of desired CPAP pressures, defined as provision of CPAP within ± one cm H2O of set pressure. RESULTS: The percentage time CPAP was within ± one cm H2O of set pressure was 66.5% for the Hudson and 71.8% for the FPH interface (p = 0.66). Oxygen saturations for the Hudson interface were in target range for a median of 97.8% of the time, and, with the FPH interface, for a median of 98.2% of the time (p = 0.76). Clinically significant differences in primary or secondary outcomes between the two groups were not detected. CONCLUSIONS: The nasal CPAP interfaces studied were equally effective in achieving desired bubble CPAP pressures and target saturations.
Keywords: Nasal continuous positive airways pressure, newborn infant, respiratory distress syndrome
DOI: 10.3233/NPM-1363612
Journal: Journal of Neonatal-Perinatal Medicine, vol. 6, no. 1, pp. 53-59, 2013
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
sales@iospress.com
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
info@iospress.nl
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office info@iospress.nl
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
china@iospress.cn
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
如果您在出版方面需要帮助或有任何建, 件至: editorial@iospress.nl