Authors: Lee, Jung Hwan | Lee, Jae Il | Kim, Dong Hwan | Nam, Kyoung Hyup | Jeon, Tae Jin | Han, In Ho*
Affiliations: Department of Neurosurgery and Medical Research Institute, Pusan National University Hospital, Busan, Korea
Correspondence: [*] Corresponding author: In Ho Han, Department of Neurosurgery and Medical Research Institute, Pusan National University Hospital, 179 Gudeok-ro, Amidong 1(il)-ga, Seo-gu, Busan, 49241, Korea. Tel.: +82 51 240 7257; Fax: +82 51 244 0282; E-mail: farlateral@hanmail.net.
Abstract: BACKGROUND: Monitoring patients’ position is important, but there have been few studies related to validation. OBJECTIVE: The objective of this study was to assess the validity of position monitoring measured using a wearable device by comparing the device’s measurements to a patient’s actual position. METHODS: We constructed a wearable device with a three-axis gyroscope and applied it to 10 patients who were unable to change their position independently. We compared the actual angle of the position and the angle transmitted from the wearable device using a Bland-Altman plot and a receiver operating characteristic curve. RESULTS: We compared the actual angle of the position and the angle transmitted from the wearable device using a Bland-Altman plot, but it was difficult to observe statistical similarity. The angles transmitted from the wearable device in the lateral and supine positions showed significant differences. The cutoff value separating the lateral and supine positions was found to be 27.1∘ (sensitivity = 100%, specificity = 99.9%). CONCLUSIONS: Through our method, the measured values from the gyroscope-based wearable device did not accurately reflect the patient’s actual position. However, the wearable device was able to distinguish the lateral position from the supine position.
Keywords: Wearable device, validation, monitoring, position change, patients
DOI: 10.3233/THC-202575
Journal: Technology and Health Care, vol. 29, no. 4, pp. 843-848, 2021
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
sales@iospress.com
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
info@iospress.nl
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office info@iospress.nl
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
china@iospress.cn
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
如果您在出版方面需要帮助或有任何建, 件至: editorial@iospress.nl