Article type: Research Article
Authors: Zens, Martina; b; * | Goldschmidtboeing, Franka | Wagner, Ferdinandb | Reising, Kilianb | Südkamp, Norbert P.b | Woias, Petera
Affiliations: [a] Department of Microsystems Engineering (IMTEK), Albert-Ludwigs-Universität Freiburg, Freiburg, Germany | [b] Department of Orthopedic Surgery and Traumatology, University Medical Center Freiburg, Freiburg, Germany
Correspondence: [*] Corresponding author: Martin Zens, Department of Microsystems Engineering (IMTEK), Albert-Ludwigs-University Freiburg, Georges-Koehler-Allee 102, 79110 Freiburg, Germany. Tel.: +49 761 203 67491; Mobile: +49 163 3374461; Fax: +49 761 203 7492; E-mail:martin.zens@imtek.uni-freiburg.de,martin.zens@me.com
Abstract: BACKGROUND: Several different surgical techniques are used in the treatment of olecranon fractures. Tension band wiring is one of the most preferred options by surgeons worldwide. The concept of this technique is to transform a tensile force into a compression force that adjoins two surfaces of a fractured bone. Currently, little is known about the resulting compression force within a fracture. OBJECTIVE: Sensor devices are needed that directly transduce the compression force into a measurement quality. This allows the comparison of different surgical techniques. Ideally the sensor devices ought to be placed in the gap between the fractured segments. METHODS: The design, development and characterization of miniaturized pressure sensors fabricated entirely from polydimethylsiloxane (PDMS) for a placement within a fracture is presented. The pressure sensors presented in this work are tested, calibrated and used in an experimental in vitro study. RESULTS: The pressure sensors are highly sensitive with an accuracy of approximately 3 kPa. A flexible fabrication process for various possible applications is described. The first in vitro study shows that using a single-twist or double-twist technique in tension band wiring of the olecranon has no significant effect on the resulting compression forces. CONCLUSIONS: The in vitro study shows the feasibility of the proposed measurement technique and the results of a first exemplary study.
Keywords: Polydimethylsiloxane (PDMS), pressure sensor, tension band wiring, compression force
DOI: 10.3233/THC-161243
Journal: Technology and Health Care, vol. 24, no. 6, pp. 909-917, 2016
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
sales@iospress.com
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
info@iospress.nl
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office info@iospress.nl
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
china@iospress.cn
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
如果您在出版方面需要帮助或有任何建, 件至: editorial@iospress.nl