Article type: Research Article
Authors: Ryf, Salomea; b | Wolber, Thomasc | Duru, Firatc | Luechinger, Rogera; *
Affiliations: [a] Institute for Biomedical Engineering, University and ETH Zurich, Switzerland | [b] Federal Office of Public Health, Berne, Switzerland | [c] Cardiovascular Center, Cardiology, University Hospital Zurich, Zurich, Switzerland
Correspondence: [*] Address for correspondence: Roger Luechinger, PhD, Institute for Biomedical Engineering, University and ETH Zürich, Gloriastrasse 35, CH-8092 Zürich, Switzerland. Tel.: +41 44 255 30 64; Fax: +41 44 255 45 06; E-mail: luechinger@biomed.ee.ethz.ch.
Abstract: Permanent magnets may interfere with the function of cardiac pacemakers and implantable cardioverter-defibrillators (ICDs). Neodymium-iron-boron (NdFeB) magnets have become widely available in recent years and are incorporated in various articles of daily life. We conducted an in-vitro study to evaluate the ability of NdFeB magnets for home and office use to cause interference with cardiac pacemakers and ICDs. The magnetic fields of ten NdFeB magnets of different size and shape were measured at increasing distances beginning from the surface until a field-strength (B-field) value of 0.5 mT was reached. Furthermore, for each magnet the distance was determined at which a sample pacemaker switched from magnet mode to normal mode. Depending on the size and remanence of individual magnets, a B-field value of 0.5 mT was found at distances ranging from 1.5 cm to 30 cm and a value of 1 mT at distances from 1 cm to 22 cm. The pacemaker behavior was influenced at distances from 1 cm to 24 cm. NdFeB magnets for home and office use may cause interference with cardiac pacemakers and ICDs at distances up to 24 centimeters. Patient education and product declarations should include information about the risk associated with these magnets.
Keywords: Pacemaker, implantable cardioverter-defibrillator, EMI, neodymium magnet
DOI: 10.3233/THC-2008-16102
Journal: Technology and Health Care, vol. 16, no. 1, pp. 13-18, 2008
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
sales@iospress.com
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
info@iospress.nl
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office info@iospress.nl
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
china@iospress.cn
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
如果您在出版方面需要帮助或有任何建, 件至: editorial@iospress.nl