Searching for just a few words should be enough to get started. If you need to make more complex queries, use the tips below to guide you.
Issue title: Papers from the Regensburg Applied Biomechanics Symposium, June 2005
Guest editors: Joachim Hammerx and Michael Nerlichy
Article type: Research Article
Authors: Füchtmeier, B.a | May, R.b | Fierlbeck, J.b | Hammer, J.b | Nerlich, M.a
Affiliations: [a] University Clinic, Department of Traumatology, Regensburg, Germany | [b] Mechanical Engineering Faculty, Laboratory for Materials Technology, University of Applied Science, Regensburg, Germany | [x] Mechanical Engineering Faculty, Laboratory for Materials Technology, University of Applied Science, Regensburg, Germany | [y] University Clinic, Department of Traumatology, Regensburg, Germany
Correspondence: [*] Address for correspondence: Dr. med. Bernd Füchtmeier, Klinikum der Universität Regensburg, Abteilung für Unfallchirurgie, Franz-Josef-Strauß Allee 11, D- 93053 Regensburg, Germany. E-mail: bernd.fuechtmeier@klinik.uni-regensburg.de.
Abstract: A new humerus nail (Sirus™) has become available for the treatment of fractures of the proximal humerus. The aim of the study was the biomechanical comparison of this implant with established systems. 12 matched pairs of human humeri were employed for testing. Three randomized groups were formed (n=4 pairs). A bending moment of 7.5 Nm and a torsional moment of 8.3 Nm were applied in a test of five loading cycles to all intact bones. Loading was consistently initiated at the distal epiphysis. The consequent deformation at the distal epiphysis was continuously recorded. Then an osteotomy with a defect of 5 mm was created to simulate an unstable subcapital humerus fracture. For paired comparison, one of each pair was stabilized with the Sirus proximal humerus nail. The other side was fixed with a reference implant. The following groups were created.: Group I: Sirus versus Proximal humerus nail (PHN) with spiral blade. Group II: Sirus versus PHILOS plate. Group III: Sirus versus 4.5 mm AO T-plate. The proximal humerus nail (Sirus) demonstrated significantly higher stiffness values than the reference implants for both bending and torsional load. The following values were recorded at a bending moment of 7.5 Nm: Sirus 14.2 mm, PHN 20.7 mm, PHILOS plate 28.1 mm, 4.5 AO T-plate 29.3 mm p<0.0012). The values at 8.3 Nm torsional moment were: Sirus 8.5°, PHN 12.3°, PHILOS plate 16.4°, 4.5 AO T-Platte 15.6° (p<0.002). The intramedullary load carriers were biomechanically superior when compared to the plating systems in the fracture model presented here. Supplementary, the Sirus Nail showed higher stiffness values than the PHN. However, the latter are gaining in importance due to the possibility of minimal invasive implantation. Whether this will be associated with functional advantages requires further clinical investigation.
Keywords: Proximal humerus fractures, biomechanical analysis, Sirus nail
DOI: 10.3233/THC-2006-144-509
Journal: Technology and Health Care, vol. 14, no. 4-5, pp. 261-270, 2006
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
sales@iospress.com
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
info@iospress.nl
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office info@iospress.nl
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
china@iospress.cn
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
如果您在出版方面需要帮助或有任何建, 件至: editorial@iospress.nl