Searching for just a few words should be enough to get started. If you need to make more complex queries, use the tips below to guide you.
Article type: Research Article
Authors: Wijaya, Sastra Kusumaa | Sumikawa, Takuyab | Saratani, Keijib | Kawazoe, Takayoshib | Oka, Hisaoc; *
Affiliations: [a] Division of Science and Technology for Intelligent, Graduate school of Natural Science and Technology, Okayama University, Japan | [b] Department of Fixed Prosthodontics and Occlusion, Osaka Dental University, Japan | [c] Faculty of Health Sciences, Okayama University, Japan
Correspondence: [*] Corresponding author: Dr. H. Oka, Department of Medical Technology, Faculty of Health Sciences, Okayama University Medical School, 2-5-1 Shikata-cho, Okayama 700-8558, Japan. Tel./Fax: +81 086 235 6884; E-mail: hoka@md.okayama-u.ac.jp.
Abstract: This study aimed at developing a dental implant movement (IM) checker for assessing quantitative dental implant mobility. The design of the instrument was based on the tooth mobility (TM) tester, which was previously developed by our group. The IM checker consists of a newly developed measuring probe which has the size of a typical dental drill so that it would be easy to measure at all regions of dental implants. The probe has a bimorph ceramics transducer for actuating an implant at constant frequency and force amplitude and for detecting acceleration response. A set of strain gauges were attached to the bimorph ceramics for detecting preload during measurement. A new digital data acquisition system was used to eliminate measurement artifacts mainly due to probe handling. The IM checker could discriminate the artificial dental implant models in the range of clinical tooth mobility M0 with variation less than 6%. The measuring time needed by five operators was less than 15 s. Accordingly, the IM checker has sufficient measuring reliability and therefore it could be introduced in dental clinics.
Keywords: tooth mobility, dental implant, mechanical mobility, manual examination
DOI: 10.3233/THC-2004-12102
Journal: Technology and Health Care, vol. 12, no. 1, pp. 11-23, 2004
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
sales@iospress.com
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
info@iospress.nl
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office info@iospress.nl
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
china@iospress.cn
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
如果您在出版方面需要帮助或有任何建, 件至: editorial@iospress.nl