Searching for just a few words should be enough to get started. If you need to make more complex queries, use the tips below to guide you.
Article type: Research Article
Authors: Zhang, Xiaomeia; b; 1 | Wang, Zhea; c; * | Yun, Xiangyua; b | Li, Mohana; c | Hu, Jinminga; b | Wang, Chengmina; b | Wei, Cunfenga; b; c; *
Affiliations: [a] Beijing Engineering Research Center of Radiographic Techniques and Equipment, Institute of High Energy Physics, Chinese Academy of Sciences, Beijing, China | [b] School of Nuclear Science and Technology, University of Chinese Academy of Sciences, Beijing, China | [c] Jinan Laboratory of Applied Nuclear Science, Jinan, China
Correspondence: [*] Corresponding author: Cunfeng Wei and Zhe Wang, Beijing Engineering Research Center of Radiographic Techniques and Equipment, Institute of High Energy Physics, Chinese Academy of Sciences, Beijing 100049, China. E-mails: weicf@ihep.ac.cn (CW), wangzhe@ihep.ac.cn (ZW).
Note: [1] ORCID ID: Xiaomei Zhang https://orcid.org/0009-0000-3332-339X
Abstract: BACKGROUND:Photon counting spectral CT is a significant direction in the development of CT technology and material identification is an important application of spectral CT. However, spectrum estimation in photon counting spectral CT is highly complex and may affect quantification accuracy of material identification. OBJECTIVE:To address the problem of energy spectrum estimation in photon-counting spectral CT, this study investigates empirical material decomposition algorithms to achieve accurate quantitative decomposition of the effective atomic number. METHODS:The spectrum is first calibrated using the empirical dual-energy calibration (EDEC) method and the effective atomic number is then quantitatively estimated based on the EDEC method. The accuracy of estimating the effective atomic number of materials under different calibration conditions is investigated by designing different calibration phantoms, and accurate quantitation is achieved using suitable calibration settings. Last, the validity of this method is verified through simulations and experimental studies. RESULTS:The results demonstrate that the error in estimating the effective atomic number is reduced to within 4% for low and medium Z materials, thereby enabling accurate material identification. CONCLUSION:The empirical dual-energy correction method can solve the problem of energy spectrum estimation in photon counting spectral CT. Accurate effective atomic number estimation can be achieved with suitable calibration.
Keywords: Photon counting detectors (PCDs), material decomposition, spectral CT, effective atomic number, polynomial fitting, calibration phantoms
DOI: 10.3233/XST-230054
Journal: Journal of X-Ray Science and Technology, vol. 31, no. 4, pp. 811-824, 2023
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
sales@iospress.com
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
info@iospress.nl
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office info@iospress.nl
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
china@iospress.cn
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
如果您在出版方面需要帮助或有任何建, 件至: editorial@iospress.nl