Searching for just a few words should be enough to get started. If you need to make more complex queries, use the tips below to guide you.
Article type: Research Article
Authors: Xu, Junjiea | Wang, Tonga | Luo, Yua | Shang, Lintaoa | Mai, Xiuyinga | Ruan, Junjiea | Pan, Xiaofenb; * | Chi, Fenga; *
Affiliations: [a] State Key Laboratory of Oncology in South China, Sun Yat-sen University Cancer Center, Guangzhou, China | [b] Department of Oncology, The Seventh Affiliated Hospital, Sun Yat-sen University, Shenzhen, China
Correspondence: [*] Corresponding authors. Feng Chi, State Key Laboratory of Oncology in South China, Sun Yat-sen University Cancer Center, 651 East Dongfeng Road, Guangzhou, China, 510060. E-mail: chifeng@sysucc.org.cn and Xiaofen Pan, Department of Oncology, The Seventh Affiliated Hospital, Sun Yat-sen University. No. 628, Zhenyuan Road, Guangming District, Shenzhen, 518100 China. E-mail: panxf8@mail.sysu.edu.cn.
Abstract: Background:There is no standardized registration frame of cone beam CT (CBCT) in intensity modulated radiotherapy (IMRT) for nasopharyngeal carcinoma (NPC). The overall registration frame that covers the whole head and neck is the most commonly used CBCT registration frame for NPC patients in IMRT. Objective:To compare the set-up errors using different registration frames of CBCT for NPC to assess the set-up errors for different region of the commonly used clinical overall registration frame. Methods:294 CBCT images of 59 NPC patients were collected. Four registration frames were used for matching. The set-up errors were obtained using an automatic matching algorithm and then compared. The expansion margin from the clinical target volume (CTV) to the planned target volume (PTV) in the four groups was also calculated. Results:The average range of the isocenter translation and rotation errors of four registration frames are 0.89∼2.41 mm and 0.49∼1.53°, respectively, which results in a significant difference in the set-up errors (p < 0.05). The set-up errors obtained from the overall frame are smaller than those obtained from the head, upper neck, and lower neck frames. The margin ranges of the overall, head, upper neck, and lower neck frames in three translation directions are 1.49∼2.39 mm, 1.92∼2.45 mm, 1.86∼3.54 mm and 3.02∼4.78 mm, respectively. The expansion margins calculated from the overall frame are not enough, especially for the lower neck. Conclusion:Set-up errors of the neck are underestimated by the overall registration frame. Thus, it is important to improve the position immobilization of the neck, especially the lower neck. The margin of the target volume of the head and neck region should be expanded separately if circumstances permit.
Keywords: Nasopharyngeal carcinoma, set-up error, registration frame, IMRT, CBCT
DOI: 10.3233/XST-230024
Journal: Journal of X-Ray Science and Technology, vol. 31, no. 5, pp. 1067-1077, 2023
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
sales@iospress.com
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
info@iospress.nl
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office info@iospress.nl
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
china@iospress.cn
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
如果您在出版方面需要帮助或有任何建, 件至: editorial@iospress.nl