Article type: Research Article
Authors: Kim, Tae-Hoona; 1 | Ryu, Jong-Hyuna; 1 | Jeong, KilHwan | Kim, Kyu Gyeoma | Jeong, Chang-Wona | Yoon, Kwon-Haa; b; * | Kim, Dae-Wonc
Affiliations: [a] Medical Convergence Research Center, Wonkwang University, Iksan, Jeonbuk, South Korea | [b] Department of Radiology, Wonkwang University School of Medicine, Iksan, Jeonbuk, South Korea | [c] Department of Neurosurgery, Wonkwang University School of Medicine, Iksan, Jeonbuk, South Korea
Correspondence: [*] Corresponding author: Kwon-Ha Yoon, Department of Radiology, Wonkwang University School of Medicine and Hospital, #895 Muwang-Ro, Iksan, Jeonlabuk-do 54538, Republic of Korea. Tel.: +82 63 859 1921; Fax: +82 63 859 1009; E-mail: khy1646@wku.ac.kr.
Note: [1] The first two authors contributed equally to this study.
Abstract: BACKGROUND:Intraoperative computed tomography (iCT) system has been developed focusing on combining the advanced imaging techniques for the best imaging modality. However, the use of iCT system in the operating rooms is limited due to the lack of flexible mobility. OBJECTIVE:This study aims to develop a mobile iCT imaging system and assess its imaging performance in a phantom study. METHODS:The mobile iCT system with mecanum omni-directional wheels has three major components namely, a rotating gantry, a slip-ring and a stationary gantry. Performance of mecanum iCT system was evaluated using the indices of signal-to-noise (SNR), contrast-to noise (CNR), and spatial resolution (MTF). Anatomical landmarks on phantom images were assessed using a 5-point scale (5 = definitely seen; 4 = probably seen; 3 = equivocal; 2 = probably not seen; and 1 = definitely not seen). RESULTS:The mecanum iCT system can be conveniently used for a whole-body scan under intraoperative conditions even in narrow operating rooms due to a smaller turning radius. The image quality of the mecanum iCT system was found to be acceptable for clinical applications (with SNR = 162.72, CNR = 134.29 and MTF = 694 μm). The diagnostic scores on the phantom images were ‘definitely seen’ value. CONCLUSIONS:The proposed mecanum iCT system achieved the improved flexible mobility and has potential to better serve as a useful imaging tool in the clinical intraoperative setting.
Keywords: Intraoperative computed tomography (iCT) system, brain, body, operating room, mobile CT
DOI: 10.3233/XST-190514
Journal: Journal of X-Ray Science and Technology, vol. 27, no. 5, pp. 907-918, 2019
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
sales@iospress.com
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
info@iospress.nl
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office info@iospress.nl
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
china@iospress.cn
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
如果您在出版方面需要帮助或有任何建, 件至: editorial@iospress.nl