Article type: Research Article
Authors: Davis, Andrew M.a; c; 1 | Pearson, Erik A.a; 1; * | Pan, Xiaochuanb | Pelizzari, Charles A.a | Al-Hallaq, Haniaa
Affiliations: [a] Department of Radiation and Cellular Oncology, University of Chicago, Chicago | [b] Department of Radiology, University of Chicago, Chicago | [c] Department of Biomedical Engineering, Johns Hopkins University, Baltimore
Correspondence: [*] Corresponding author: Erik A. Pearson, Department of Radiation and Cellular Oncology, University of Chicago, 5758 S Maryland Ave, Chicago. E-mail: epearson@uchicago.edu.
Note: [1] These authors contributed equally.
Abstract: BACKGROUND:Some patients cannot be imaged with cone-beam CT for image-guided radiation therapy because their size, pose, or fixation devices cause collisions with the machine. OBJECTIVE:To investigate imaging trajectories that avoid such collisions by using virtual isocenter and variable magnification during acquisition while yielding comparable image quality. METHODS:The machine components most likely to collide are the gantry and kV detector. A virtual isocenter trajectory continuously moves the patient during gantry rotation to maintain an increased separation between the two. With dynamic magnification, the kV detector is dynamically moved to increase clearance for an angular range around the potential collision point while acquiring sufficient data to maintain the field-of-view. Both strategies were used independently and jointly with the resultant image quality evaluated against the standard circular acquisition. RESULTS:Collision avoiding trajectories show comparable contrast and resolution to standard techniques. For an anthropomorphic phantom, the RMSE is <7×10- 4, multi-scale structural similarity index is >0.97, and visual image fidelity is >0.96 for all trajectories when compared to a standard circular scan. CONCLUSIONS:The proposed trajectories avoid machine-patient collisions while providing comparable image quality to the current standard thereby enabling CBCT imaging for patients that could not otherwise be scanned.
Keywords: IGRT, CBCT, collision avoidance, noncircular trajectories, variable magnification, virtual isocenter
DOI: 10.3233/XST-180401
Journal: Journal of X-Ray Science and Technology, vol. 27, no. 1, pp. 1-16, 2019
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
sales@iospress.com
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
info@iospress.nl
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office info@iospress.nl
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
china@iospress.cn
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
如果您在出版方面需要帮助或有任何建, 件至: editorial@iospress.nl