Abstract: PURPOSE: To develop a novel scatter correction method without additional patient dose for dual-energy digital mammography (DEDM) to reduce scatter's impacts and enhance microcalcification detectability in dual-energy X-ray subtraction image. METHODS: Combining scatter radiation is lower spatial frequency component and calcifications are sparsely distributed in digital mammogram, we develop a new scatter correction strategy. First, an adaptive sampling scheme is presented to find possible noncalcification (zero calcification) pixels. Then the maximum likelihood expectation maximization (MLEM) algorithm is applied to evaluate initial scatter surface. The accurate scatter radiation of sampling pixels is obtained by solving dual-energy computational formula with zero calcification constraint and scatter surface constraint. RESULTS: After scatter correction, the scatter-to-primary ratio (SPR) of wedge phantom is reduced from ∼36.0% to ∼3.1% for low-energy (LE) image and ∼29.6% to ∼0.6% for high-energy (HE) image. For step phantom, the SPR is reduced from ∼42.1% and ∼30.3% to ∼3.9% and ∼0.9% for LE and HE image, respectively. The calcification contrast-to-noise ratio is improved by two orders of magnitudes in calcification images. CONCLUSIONS: The proposed method shows an excellent performance on scatter reduction and calcification detection. Compared with hardware based scatter correction strategy, our method need no extra exposure and is easy to implementation.
Keywords: Scatter correction, dual-energy mammography, calcification, monte carlo
DOI: 10.3233/XST-140452
Journal: Journal of X-Ray Science and Technology, vol. 22, no. 5, pp. 653-671, 2014
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
sales@iospress.com
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
info@iospress.nl
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office info@iospress.nl
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
china@iospress.cn
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
如果您在出版方面需要帮助或有任何建, 件至: editorial@iospress.nl