Article type: Research Article
Authors: Zhang, Jie | Zhang, Da
Affiliations: Division of Medical Physics, Department of Radiology, University of Kentucky, Lexington, KY, USA | Division of Diagnostic Imaging Physics and Webster Center for Advanced Research and Education in Radiation, Department of Radiology, Massachusetts General Hospital, Boston, MA, USA
Note: [] Corresponding author: Jie Zhang, Division of Medical Physics, Department of Radiology, University of Kentucky, 800 Rose Street, HX311A, Lexington, KY 40536, USA. Tel.: +1 859 323 2146; Fax: +1 859 257 4457; E-mail: jnzh222@uky.edu
Abstract: Image performance of a cone beam O-arm system with the use of a Varian flat panel detector (PaxScan 4030CB) was evaluated. Modulation Transfer Function (MTF), Noise Power Spectrum (NPS), and tube output characteristics were measured. Image quality measurements were performed with a CATPHAN 424 phantom using a 120 kVp beam filtered with 1 mm Al and 0.1 mm Cu and three preset protocols at different exposure levels. The CPT445 Module was used to measure MTF. Water phantom images were used to calculate NPS. Then the NPS was normalized by dividing the square of the mean signal. The Photon fluence at each exposure (q) was estimated using the SpekCalc. The MTF indicates that the reconstruction filter of the O-arm is overenhancing, meaning that it yields values greater than unity. The system has a 10% MTF at 0.45 mm. The NPS shows the low frequency structured noise. The evaluation of image performance of the O-arm imaging system helps its future improvements and clinical applications.
Keywords: O-arm, cone-beam computed tomography (CT), modulation transfer function (MTF), noise power spectrum (NPS), performance, flat panel detector (FPD)
DOI: 10.3233/XST-130385
Journal: Journal of X-Ray Science and Technology, vol. 21, no. 3, pp. 373-380, 2013
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
sales@iospress.com
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
info@iospress.nl
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office info@iospress.nl
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
china@iospress.cn
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
如果您在出版方面需要帮助或有任何建, 件至: editorial@iospress.nl