Article type: Research Article
Authors: Harrell, M.D.a; * | Dobson, N.R.b | Olsen, C.c | Ahmed, A.c | Hunt, C.E.b
Affiliations: [a] Neonatal, Perinatal Medicine Fellowship, National Capital Consortium, Bethesda, MD, USA | [b] Department of Pediatrics, Uniformed Services University, Bethesda, MD, USA | [c] Department of Preventive Medicine and Biostatistics, Uniformed Services University, Bethesda, MD, USA
Correspondence: [*] Address for correspondence: Mathew D. Harrell DO, Department of Pediatrics, Tripler Army Medical Center, 1 Jarrett White Road, Honolulu, HI, 96818, USA. Tel.: +1 937 371 9746; E-mail: mdharr06@gmail.com.
Abstract: BACKGROUND:To compare oxygen saturation (SpO2) and heart rate (HR) recorded by a reference wired pulse oximeter to a wireless pulse oximeter in inpatient neonates. METHODS:Term infants born≥37 + 0 weeks and preterm infants born≤35 + 0 weeks gestation were enrolled and time-matched data pairs were obtained. The primary outcome was intraclass correlation coefficient and r-values between the two oximeters for heart rate and oxygen saturation. RESULTS:Thirty term and 20 preterm neonates were enrolled. There was a high degree of correlation between the two oximeters for HR (r = 0.926) among all 50 infants, and excellent interclass correlation (ICC = 0.961), but there were no bradycardia episodes in either term or preterm infants. There was a lesser degree of correlation for SpO2 values in the term and preterm groups (r = 0.242; 0.521, respectively) along with moderate interclass correlation (ICC = 0.719) but few episodes of hypoxemia≤90% occurred in enrolled subjects. CONCLUSIONS:There were no significant differences between the wireless and reference wired oximeters for assessing HR. There was less correlation between the two oximeters for monitoring SpO2 in both the term and preterm group. Wireless pulse oximetry may have practical advantages for use in inpatient neonates, but additional studies are needed that include bradycardia and desaturation events to delineate this question.
Keywords: Oximetry, neonate, wireless pulse oximeter, wired pulse oximeter
DOI: 10.3233/NPM-210836
Journal: Journal of Neonatal-Perinatal Medicine, vol. 15, no. 2, pp. 283-289, 2022
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
sales@iospress.com
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
info@iospress.nl
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office info@iospress.nl
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
china@iospress.cn
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
如果您在出版方面需要帮助或有任何建, 件至: editorial@iospress.nl