Article type: Research Article
Authors: McClure, C. | Jang, S. Young | Fairchild, K.*
Affiliations: Department of Pediatrics, University of Virginia School of Medicine, Charlottesville, VA, USA
Correspondence: [*] Address for correspondence: Karen Fairchild, Department of Pediatrics, P.O. Box 800386, University of Virginia, Charlottesville, VA 22908, USA. Tel.: +1 434 924 5428; Fax: +1 434 924 2816; E-mail: kdf2n@virginia.edu.
Abstract: BACKGROUND: Alarm overload is a significant concern in the Neonatal Intensive Care Unit (NICU). Selecting a longer oxygen saturation (SpO2) averaging time will reduce the number of alarms but may mask fluctuations in oxygenation. OBJECTIVE: Characterize bedside monitor alarms in the NICU and estimate the impact of longer SpO2 averaging time and alarm delay. METHODS: All bedside monitor alarms were analyzed over a 12-month period in the University of Virginia NICU, using the default averaging time (8 seconds) and SpO2 alarm limits set at 88–95% for infants on supplemental oxygen. In 10 VLBW infants, SpO2 averaging time was lowered to 2 seconds for 24 hours and events of SpO2 out of the target range were estimated with 2-, 8-, or 16-second averaging time, with and without a 15-second alarm delay. RESULTS: There were 3,263,590 alarms in the unit over 12 months. Low or high SpO2 alarms constituted 79% of the total, and 55% of these events lasted <15 seconds. In 10 infants we estimated that increasing SpO2 averaging time from 2 to 16 seconds would have led to 53% fewer SpO2 alarms but the mean duration of alarms would have been 2.15-fold longer. Adding a 15-second alarm delay to 2-second SpO2 averaging in this analysis decreased SpO2 alarms by 67%. CONCLUSION: Longer SpO2 averaging times mask the number and severity of events of aberrant oxygenation in preterm infants without decreasing total alarm time. Incorporating an alarm delay with shorter SpO2 averaging times can reduce alarm number and duration, and allow more accurate assessment of oxygenation, which may be important for research into consequences of aberrant oxygenation in this vulnerable population.
Keywords: Alarm, pulse oximetry, oxygen saturation, preterm infant, averaging, hypoxemia, very low birthweight
DOI: 10.3233/NPM-16162
Journal: Journal of Neonatal-Perinatal Medicine, vol. 9, no. 4, pp. 357-362, 2016
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
sales@iospress.com
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
info@iospress.nl
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office info@iospress.nl
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
china@iospress.cn
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
如果您在出版方面需要帮助或有任何建, 件至: editorial@iospress.nl