Affiliations: Institute of Atmospheric Physics, Chinese Academy of
Sciences, Beijing 100029, China | Graduate School of the Chinese Academy of Sciences,
Beijing 100049, China
Abstract: Combining improved injector, gas line and valve-driving modules, a
GC equipped with FID and ECD, could simultaneously measure
CH_4, CO_2 and N_2O in
an air sample within 4 min. Test results showed that the system has high
sensitivity, resolution and precision; the linear response range of the system
meets the requirement of in situ flux measurements. Thus, the system is
suitable for monitoring fluxes of main greenhouse gases in terrestrial
ecosystem since it is easy to use, efficacious, stable and reliable to collect
data.
Keywords: GC, valve, emission of greenhouse gases, ecosystem