Article type: Research Article
Authors: Huerta Ojeda, Álvaroa; * | Beltrán-Inostroza, Rodrigoa; b | Yeomans-Cabrera, María-Mercedesc | Torres-Banduc, Maximilianoa; d
Affiliations: [a] Núcleo de Investigación en Salud, Actividad Física y Deporte ISAFYD, Universidad de Las Américas, Viña del Mar, Chile | [b] Escuela de Kinesiología, Facultad de Ciencias de la Salud, Universidad de Las Américas, Viña Del Mar, Chile | [c] Facultad de Salud y Ciencias Sociales, Universidad de Las Américas, Viña del Mar, Chile | [d] Escuela de Kinesiología, Facultad de Ciencias de la Salud, Universidad de Las Américas, Viña Del Mar, Chile
Correspondence: [*] Corresponding author: Álvaro Huerta Ojeda, Avenida 7 Norte 1348, Viña del Mar, 2531098, Chile. E-mail: achuertao@yahoo.es.
Abstract: BACKGROUND: Isometric maximal voluntary force (IMVF) is essential for individuals’ health and physical performance. Therefore, there is a need for valid and reliable devices to assess IMVF. OBJECTIVE: To determine the validity and reliability of the Valkyria Trainer Balance® portable force platform in its isometric mode. METHODS: Fifty-eight physically healthy individuals (30 men and 28 women) participated in the study. A repeated measures design was used to compare the inter-day test-retest reliability of peak force. The validity of the Valkyria Trainer Balance® force platform was determined by comparing the peak force with the ArtOficio® force platform. The analysis consisted of the intraclass correlation coefficient (ICC), standard error of measurement (SEM), and coefficient of variation (CV). A CV ⩽ 10% and ICC ⩾ 0.80 were considered acceptable reliability, while a ⩽ 5% and ICC ⩾ 0.90 were regarded as high reliability. RESULTS: CV the peak force showed high test-retest inter-day reliability (CV = 4.3% and ICC = 0.99). When comparing both force platforms, there was a 1.1% difference between the two devices. CONCLUSIONS: The results of this study demonstrate that the Valkyria Trainer Balance® force platform is valid and reliable for assessing IMVF in physically healthy individuals.
Keywords: Validation study, reproducibility of results, equipment and supplies, muscle contraction
DOI: 10.3233/IES-240022
Journal: Isokinetics and Exercise Science, vol. 32, no. 4, pp. 337-348, 2024
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
sales@iospress.com
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
info@iospress.nl
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office info@iospress.nl
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
china@iospress.cn
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
如果您在出版方面需要帮助或有任何建, 件至: editorial@iospress.nl