Searching for just a few words should be enough to get started. If you need to make more complex queries, use the tips below to guide you.
Article type: Research Article
Authors: Reinhart, W.H. | Häusler, K. | Schaller, P. | Erhart, S. | Stetter, M. | Dual, J. | Sayir, M.
Affiliations: Internal Medicine, Kantonsspital, CH 7000 Chur, Switzerland | Institute of Mechanics, ETH Zürich, CH 8092 Zürich, Switzerland
Abstract: A newly designed type of oscillating viscometer was tested for blood viscosity measurements. The viscometer consists of a probe (either a tube or a rod) oscillating at a resonance frequency with amplitudes in the micro‐ and nanometer range. The torsional oscillations are dampened by fluids flowing through the tube or surrounding the rod. The degree of damping depends on the viscosity of the fluid, which allows to measure viscosity. Data obtained with these instruments were compared with those obtained with a conventional Couette viscometer. An increase of erythrocyte aggregation by the addition of dextran 70 in vitro led to the expected increase of viscosity in the Couette viscometer; in the oscillating tube viscometer, however, it remained unchanged, which may be explained by a decreased erythrocyte concentration near the tube wall due to increased aggregation and flow of erythrocytes in the tube center. In ex vivo experiments on blood flowing without anticoagulant directly through the tube viscometer an inverse correlation between viscosity and fibrinogen concentration was found. This is in contrast to actual knowledge and may indicate that high fibrinogen levels have a beneficial rheological effect at the tube or vessel wall. Our data suggest that the new oscillating tube viscometer is an interesting tool, which may contribute to a more comprehensive understanding of blood flow.
Keywords: Aggregation, blood, erythrocyte, fibrinogen, viscosity
Journal: Clinical Hemorheology and Microcirculation, vol. 18, no. 1, pp. 59-65, 1998
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
sales@iospress.com
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
info@iospress.nl
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office info@iospress.nl
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
china@iospress.cn
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
如果您在出版方面需要帮助或有任何建, 件至: editorial@iospress.nl