Article type: Research Article
Authors: Greiner, Barbaraa; * | Kaiser, Ulrichb | Maurer, Katharinaa | Stroszczynski, Christiana | Jung, Ernst Michaela
Affiliations: [a] Institute of Diagnostic Radiology and Interdisciplinary Ultrasound, University Hospital Regensburg, Regensburg, Germany | [b] Medical Clinic and Polyclinic III, University Hospital Regensburg, Regensburg, Germany
Correspondence: [*] Corresponding author: Dr. med. univ. Barbara Greiner, Institute of Diagnostic Radiology and Interdisciplinary Ultrasound, University Hospital Regensburg, Franz-Josef-Strauß Allee 11, Regensburg 93053 Germany. Tel.: +49 941 944 17492; E-mail: barbara.greiner@klinik.uni-regensburg.de.
Abstract: BACKGROUND: Rapid evolution of ultrasound technology has allowed widespread use of handheld ultrasound devices (HHUDs) for many possible applications. Along with the adult population, the use of HHUDs for Point of Care Ultrasound (POCUS) in pediatric medicine has been increasing over the last few years. However, pediatric-specific literature is still scarce on mobile vascular ultrasound. OBJECTIVE: To evaluate diagnostic capabilities of Vscan Air™ in comparison with high-end ultrasound for the assessment of the internal jugular vein in children and adolescents. METHODS: 42 Internal Jugular Veins (IJVs) of 21 pediatric patients were scanned by an experienced examiner using a WLAN-supported handheld ultrasound device (Vscan Air™) and high-end cart-based ultrasound (LOGIQ E9) as reference. B-Mode and Color-coded Doppler (CCDS) were performed and compared. Image quality was assessed using a score of 0 to 5 and statistically analyzed. Results were interpreted independently by two readers in consensus. RESULTS: 21 patients (2–17 years; mean 11,00±4,5 years; female n = 11, male n = 10) were examined. The rating score never dropped below 3 for both devices. The median score evaluation of B-Mode and CCDS for the high-end device was 5.00, of Vscan Air™ 5.00 for B-Mode and 4.00 for CCDS. A significant difference was shown between the two devices in the evaluation of CCDS. CONCLUSIONS: Vscan Air™ ultrasound device allows sufficient assessability of the IJV in pediatric patients, opening up new possibilities for fast and mobile POCUS of cervical veins and potential guidance of central venous catheter placement.
Keywords: Handheld ultrasound, Vscan Air, POCUS, internal jugular vein, pediatric
DOI: 10.3233/CH-231981
Journal: Clinical Hemorheology and Microcirculation, vol. 86, no. 4, pp. 441-449, 2024
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
sales@iospress.com
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
info@iospress.nl
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office info@iospress.nl
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
china@iospress.cn
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
如果您在出版方面需要帮助或有任何建, 件至: editorial@iospress.nl