Article type: Research Article
Authors: Jin, Lina | Tong, Lan-Yueb | Sha, Leib | Cao, Meng-Mengb | Shen, Cui-Qinb | Du, Lian-Fangc | Li, Zhao-Junb; c; *
Affiliations: [a] Department of Ultrasound, Jiading District Central Hospital Affiliated Shanghai University of Medicine & Health Sciences, Shanghai, China | [b] Department of Ultrasound, Shanghai General Hospital Jiading Branch, Shanghai Jiaotong University School of Medicine, Shanghai, China | [c] Department of Ultrasound, Shanghai General Hospital, Shanghai Jiaotong University School of Medicine, Shanghai, China
Correspondence: [*] Corresponding author: Zhaojun Li, Department of Ultrasound, Shanghai General Hospital, Shanghai Jiaotong University School of Medicine, 100 Haining Road, Hongkou District, Shanghai 200080, China. Tel.: +86 18964938312; Fax: +86 021 37798276; E-mail: lzj_1975@shsmu.edu.cn. ORCID:0000-0001-6459-3667.
Abstract: AIM:Carotid ultrasound is a key tool for the diagnosis and evaluation of cardio disease, and the measurement of carotid intima-media thickness (CIMT) and hemodynamic parameters is of paramount importance for the imaging method. The aim of this study was to evaluate the feasibility and accuracy of handheld ultrasound devices for measuring carotid parameters. METHODS:We performed a carotid ultrasound on 25 participants using a handheld ultrasound device and a conventional ultrasound machine. For each participant, max and mean CIMT of common carotid artery (CCA) and peak systolic velocity (PSV), end diastolic velocity (EDV) and resistive index (RI) of CCA, bilateral external carotid artery (ECA), internal carotid artery (ICA) and the vertebral artery were measured. Agreement and repeatability were evaluated by linear regression and Bland-Altman analysis. RESULTS:We found a good repeatability and consistent of handheld ultrasound device in measuring mean CIMT (r = 0.68, P < 0.01). Furthermore, there was a moderate to good agreement between handheld and conventional ultrasound systems in measuring max IMT, mean IMT, PSV, EDV and RI of CCA (0.73, 0.79, 0.52, 0.58 and 0.84, respectively). CONCLUSION:Handheld ultrasound devices were able to provide carotid IMT and hemodynamic parameters measurements similar to those of conventional ultrasound. Such capabilities of handheld ultrasound devices might be useful for the primary assessment of carotid in clinical work.
Keywords: Handheld ultrasound scanner, carotid artery, carotid intima-media thickness, Ultrasound
DOI: 10.3233/CH-221521
Journal: Clinical Hemorheology and Microcirculation, vol. 82, no. 3, pp. 255-263, 2022
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
sales@iospress.com
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
info@iospress.nl
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office info@iospress.nl
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
china@iospress.cn
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
如果您在出版方面需要帮助或有任何建, 件至: editorial@iospress.nl