Searching for just a few words should be enough to get started. If you need to make more complex queries, use the tips below to guide you.
Issue title: Frontiers in Biomedical Engineering and Biotechnology – Proceedings of the 2nd International Conference on Biomedical Engineering and Biotechnology, 11–13 October 2013, Wuhan, China
Article type: Research Article
Authors: Wang, Xian-Kang | Ye, Jin-Duo | Gu, Fu-Shun | Wang, Ai-guo | Zhang, Chun-Qiu; | Tian, Qian-Qian | Li, Xue | Dong, Li-Min
Affiliations: Tianjin Key Laboratory for Control Theory & Applications in Complicated Industry Systems, School of Mechanical Engineering, Tianjin University of Technology, Tianjin 300384, China | Tianjin University of Chinese Medicine, Second Affiliated Hospital, Tianjin 300150, China
Note: [] Corresponding author: Chun-Qiu Zhang. E-mail: zhang_chunqiu@126.com
Abstract: In this paper, the single arm external fixation of intertrochanteric fracture healing process after surgery was simulated to obtain a postoperative fracture healing and stress distribution in the external fixator. Firstly CT images of intertrochanteric fracture are reconstructed into the femur solid model. Then based, the external fixator is installed on the model, which lastly formed a finite element model of unilateral external fixation for intertrochanteric fracture. The calculated results show: during the beginning of the fracture healing, there is much higher stress in both screws and femur in the model with solid screws than that in the model with hollow screw. The stress of the femur in the model with hollow screw is more evenly. During the middle time of Fracture healing, stress in the femoral head significantly decreases. And the stress at fracture site gradually increased with the healing occurrence. According to the results, the authors designed hollow screws to use external fixation surgery. Surgery confirmed that the use of hollow screws in fractures treatment can satisfy the strength requirements, and can effectively reduce operative time, less patient suffering. The research for external fixation can provide a reference, and promote the use of external fixation hollow screws.
Keywords: single arm external fixation, the femoral intertrochanteric fracture, stress distribution, numerical simulation, surgery scheme
DOI: 10.3233/BME-130850
Journal: Bio-Medical Materials and Engineering, vol. 24, no. 1, pp. 625-632, 2014
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
sales@iospress.com
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
info@iospress.nl
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office info@iospress.nl
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
china@iospress.cn
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
如果您在出版方面需要帮助或有任何建, 件至: editorial@iospress.nl