Article type: Research Article
Authors: Besong, A.A. | Hailey, J.L. | Ingham, E. | Stone, M. | Wroblewski, B.M. | Fisher, J.
Affiliations: Department of Mechanical Engineering, University of Leeds, LS2 9JT, UK | Department of Microbiology, University of Leeds, LS2 9JT, UK | Department of Orthopaedics, Leeds General Infirmary, UK | Centre for Hip Surgery, Wrightington Hospital, Wigan, UK
Abstract: Damage to polished femoral heads in vivo can cause increased wear of acetabular cups. Oxidation and ageing after sterilisation by gamma irradiation in air, can also change the mechanical properties and wear resistance of ultra high molecular weight polyethylene (UHMWPE). This study investigated the combined effect of these changes in material properties on the wear of UHMWPE for different counterface roughnesses, representative of new femoral heads and those damaged in vivo. Wear rates were studied on a tri-pin-on-disc tribometer in a protein-containing solution. A comparison was made of the wear, using three different counterface roughnesses, of specimens that were manufactured from polyethylene acetabular cups of different shelf ages (3–120 months) after gamma irradiation in air but never implanted. These were compared to the wear of control specimens that were manufactured from cups that had not been sterilised. The wear surfaces were tested 1 mm below the initial articulating surface of the cup, the position of high degradation. The wear rate of UHMWPE which had been sterilised by gamma irradiation in air was shown to increase significantly with ageing time on the shelf for all counterface conditions. The wear rate of all materials increased markedly as the counterface roughness increased, but to different extents depending on the age of the material. The combined effect of ageing and increase in counterface roughness had a dramatic effect (as high as 2000 fold increase) on the wear rate. Both ageing of the polymer and damage to the femoral head have been cited as causing increased wear in vivo. The results of this study demonstrate that these variables can act synergistically to markedly effect UHMWPE wear rate.
DOI: 10.3233/BME-1997-7105
Journal: Bio-Medical Materials and Engineering, vol. 7, no. 1, pp. 59-65, 1997
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
sales@iospress.com
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
info@iospress.nl
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office info@iospress.nl
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
china@iospress.cn
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
如果您在出版方面需要帮助或有任何建, 件至: editorial@iospress.nl