Article type: Research Article
Authors: Ozeki, K.; | Mishima, A. | Yuhta, T. | Fukui, Y. | Aoki, H.
Affiliations: Frontier Research and Development Center, Tokyo Denki University, Ishizaka, Hatoyama, Hiki, Saitama 350‐0394, Japan | Iwamizawa 7‐jo Dental Clinic, 7‐jo, 4‐chome, Iwanizawa, Hokkaido 068‐0027, Japan | Applied Systems Engineering, Graduate School of Science and Engineering, Tokyo Denki University, Ishizaka, Hatoyama, Hiki, Saitama 350‐0394, Japan
Note: [] Corresponding author. Tel.: +81 (0)49 296 2911(3752); Fax: +81 (0)49 296 2925; E‐mail: ozeki@bme.f.dendai.ac.jp.
Abstract: Hydroxyapatite (HA) was coated onto titanium substrates using radio frequency sputtering, and the coated HA films were crystallized in an autoclave at 110°C using a low temperature hydrothermal method. The crystallite size, the Ca/P ratio, and the surface of the films were observed using XRD, EDS, and SEM, respectively. An immersion test was carried out in physiological saline solution, and the film‐to‐substrate adhesion strength was measured using a pull‐out test. From the XRD patterns, a sputtered film subjected to the hydrothermal treatment had crystallized after 24 h, and the crystallite size increased from 38.0±8.7 to 81.4±19.1 nm. In immersion tests, an as‐sputtered film completely dissolved after 1 d, whereas a sputtered film subjected to the hydrothermal treatment survived to 18.7±5.8% of the initial film thickness after four weeks. In pull‐out tests, the adhesion strength of the sputtered film to the substrate increased from 1.9±0.2 to 5.3±1.6 MPa after the hydrothermal treatment. A sputtered film subjected to the hydrothermal treatment and a plasma‐sprayed coating on titanium columns were implanted in the diaphysis of the femora of six adult dogs, and a pull‐out test was carried out after two, four, and 12 weeks. The sputtered film showed higher bone bonding strength than the plasma‐sprayed coating at any period.
Keywords: Sputtering, hydroxyapatite, crystallization, hydrothermal, bone bonding strength
Journal: Bio-Medical Materials and Engineering, vol. 13, no. 4, pp. 451-463, 2003
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
sales@iospress.com
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
info@iospress.nl
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office info@iospress.nl
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
china@iospress.cn
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
如果您在出版方面需要帮助或有任何建, 件至: editorial@iospress.nl