Searching for just a few words should be enough to get started. If you need to make more complex queries, use the tips below to guide you.
Article type: Research Article
Authors: Kim, Sunmina | Yonemitsu, Ikuoa; | Takemura, Hiroshib | Shimoda, Kuniob | Suga, Kazuhiroc | Soga, Koheid | Ono, Takashia | Uo, Motohiroe
Affiliations: [a] Department of Orthodontic Science, Graduate School of Medical and Dental Sciences, Tokyo Medical and Dental University (TMDU), Tokyo, Japan | [b] Department of Mechanical Engineering, Faculty of Science and Technology, Tokyo University of Science, Chiba, Japan | [c] Department of Mechanical Engineering, Faculty of Engineering, Kogakuin University, Tokyo, Japan | [d] Department of Materials Science and Technology, Faculty of Industrial Science and Technology, Tokyo University of Science, Tokyo, Japan | [e] Department of Advanced Biomaterials, Graduate School of Medical and Dental Sciences, Tokyo Medical and Dental University (TMDU), Tokyo, Japan
Correspondence: [*] Corresponding author: Ikuo Yonemitsu, 1-5-45 Yushima, Bunkyo-ku, Tokyo 113-8510, Japan. Tel.: +81 3 5803 5530; Fax: +81 3 5803 5530; E-mail: yoneman.orts@tmd.ac.jp
Abstract: BACKGROUND:Most orthodontic cases consist of varying degrees of crowding. To manage crowded dentitions, nickel-titanium archwires with various ligation methods are often used. OBJECTIVE:We aimed to investigate the effect of different ligation methods with respect to force and moment and suggest the efficient ligation method for treating rotation and displacement simultaneously. METHODS:We built a model that simulated the three anterior teeth of the maxilla. The teeth on the two ends were fixed, and the middle tooth was set in several different positions by manipulating the amount of displacement in bucco-lingual direction and rotation angle. The measurements were taken with three different ligation methods of self-ligation (SL), elastomeric o-ring ligation on both side wings (EB), and on one side wings (EO). RESULTS:The magnitude of linguo-buccal force exceeded the standard optimal force in each condition examined and was significantly larger in EB than in other ligation methods. Moreover, the magnitude of moment generation with SL was suitable in the 0.0 mm linguo-version, whereas it was suitable with EO in the linguo-version ranging 1.0–3.0 mm. CONCLUSIONS:The ligation method significantly affected the force and moment. SL and EO are recommended in dentitions with light and deep lingual displacements, respectively.
Keywords: Six-axis sensor, orthodontics, force system, moment, orthodontic simulator
DOI: 10.3233/BME-211288
Journal: Bio-Medical Materials and Engineering, vol. 33, no. 6, pp. 453-463, 2022
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
sales@iospress.com
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
info@iospress.nl
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office info@iospress.nl
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
china@iospress.cn
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
如果您在出版方面需要帮助或有任何建, 件至: editorial@iospress.nl