Searching for just a few words should be enough to get started. If you need to make more complex queries, use the tips below to guide you.
Article type: Research Article
Authors: Mataliotakis, George I.a; * | Tsouknidas, Alexanderb | Panteliou, Sofiac | Vekris, Marios D.a | Mitsionis, Grigorios I.a | Agathopoulos, Simeond | Beris, Alexander E.a
Affiliations: [a] Department of Orthopaedic Surgery, University of Ioannina, Ioannina, Greece | [b] Department of Mechanical Engineering, Aristotle University of Thessaloniki, Thessaloniki, Greece | [c] Department of Mechanical Engineering and Aeronautics, University of Patras, Patras, Greece | [d] Department of Materials Science and Engineering, University of Ioannina, Ioannina, Greece
Correspondence: [*] Address for correspondence: George I. Mataliotakis, 5 Rehearsal Rooms, 115-119 Westgate Road, Newcastle upon Tyne NE1 4BD, UK. Tel.: +44 7454650853; E-mail: george.mataliotakis@gmail.com.
Abstract: BACKGROUND: The optimum fixation device for the critical size bone defect is not established yet. OBJECTIVE: A reliable, feasible and low-cost fixation device for the long-term maintenance of a critical bone defect. METHODS: A custom-made plate made of poly-methyl-methacrylate was used for the fixation of a critical defect of rats’ femurs. The screws were securely fixing both on the plate and the bone. A three point bending test, aimed to resemble the in vivo loading pattern, a Finite Element Analysis and a 24-week in vivo monitoring of the integrity of the plate fixation were utilized. RESULTS: The plate has linear and reproducible behavior. It presents no discontinuities in the stress field of the fixation. Its properties are attributed to the material and the locking principle. It fails beyond the level of magnitude of the normal ambulatory loads. In vivo, 100% of the plates maintained the bone defect intact up to 12 weeks and 85% of them at 24 weeks. CONCLUSION: This novel locking plate shows optimal biomechanical performance and reliability with high long-term in vivo survival rate. It is fully implantable, inexpensive and easily manufactured. It can be qualified for long term critical defect fixation in bone regeneration studies.
Keywords: Bone tissue engineering, segmental bone defects, animal models, plate, critical size defect, fixation device
DOI: 10.3233/BME-151540
Journal: Bio-Medical Materials and Engineering, vol. 25, no. 4, pp. 335-346, 2015
IOS Press, Inc.
6751 Tepper Drive
Clifton, VA 20124
USA
Tel: +1 703 830 6300
Fax: +1 703 830 2300
sales@iospress.com
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
IOS Press
Nieuwe Hemweg 6B
1013 BG Amsterdam
The Netherlands
Tel: +31 20 688 3355
Fax: +31 20 687 0091
info@iospress.nl
For editorial issues, permissions, book requests, submissions and proceedings, contact the Amsterdam office info@iospress.nl
Inspirees International (China Office)
Ciyunsi Beili 207(CapitaLand), Bld 1, 7-901
100025, Beijing
China
Free service line: 400 661 8717
Fax: +86 10 8446 7947
china@iospress.cn
For editorial issues, like the status of your submitted paper or proposals, write to editorial@iospress.nl
如果您在出版方面需要帮助或有任何建, 件至: editorial@iospress.nl